Manual

27
GLDxxxfff:Requests the records of the “xxx” lot number. The records are
sent in frames of 10 records; “fff” is the frame number (01 first
frame). (Example: Lot 13 has 53 records. The records will be
sent in 6 frame, 5 with10 records and 1 with 3 records.)
Command Parameters:
xxx - Lot number (eq: for lot number 1 xxx = 001)
fff - requested frame number - first frame is labeled 01
Notes: “Err3” is sent if the Log on demand is empty.
“Err4” is sent if the requested set parameter is not available.
“Err5” is sent if the command argument is wrong.
“Err6” is sent if the requested range is not available.
“Err7” is sent if the instrument is logging.
“Err8” is sent if the instrument is not in measurement mode.
Invalid commands will be ignored.