CHR xx Change the instrument range according with the parameter
value (xx):
xx=06 % range
xx=07 ppm range
The instrument will answer for these commands with:
<STX> <answer
>
<ETX>
where: <STX> is 02 ASCII code character (start of text)
<ETX> is 03 ASCII code character (end of text)
<answer>:
<ACK> is 06 ASCII code character (recognized command)
<NAK> is 21 ASCII code character (unrecognized command)
<CAN> is 24 ASCII code character (corrupted command)
COMMANDS REQUIRING AN ANSWER
The instrument will answer for these commands with:
<STX> <answer> <checksum> <ETX>
where the checksum is the bytes sum of the answer string sent as 2 ASCII
characters.
All the answer messages are with ASCII characters.
RAS Causes the instrument to send a complete set of readings.
The answer string contains:
Meter mode (2 chars):
06 - % range
07 - ppm range
Meter status (2 chars of status byte): represents a 8
bit hexadecimal encoding.
0x10 - temperature probe is connected
0x01 - new GLP data available
0x02 - new SETUP parameter
Reading status (1 char): R - in range, O - over
range, U - under range.
The reading (corresponding to the selected range) -
7 ASCII chars, including sign and decimal point.
Temperature reading - 7 ASCII chars, with sign and
two decimal points, always in ºC.