ZERO CALIBRATION
Submerse the probe into HI 7040
zero oxygen
solution and stir gently for 2-3 minutes.
Press CAL. The “~” and “ ” tags will blink on the
LCD until the reading is stable.
When the reading is stable, “CFM” starts blinking.
Press CFM to confirm the “0.0%” DO calibration.
If the reading is within the limits (±15% f.s.), the
meter stores the value (and adjusts the slope point).
Press CAL. The instrument will return to measurement
mode and will memorize the zero calibration data.
For a two-point calibration do not press CAL and
follow the procedure below.
SLOPE CALIBRATION
It is suggested to perform the slope calibration in air.
Rinse the probe in clean water to remove any residual
zero oxygen solution.
Note: If you did not perform the zero calibration procedure, press CAL
and then the ARROW keys to select the 100% DO calibration
point.
Dry the probe tip and allow a few seconds for the LCD reading to
stabilize. The “
~~
~~
~” and “ ” tags will blink until the reading is
stable.
When the reading is stable, “CFM” tag starts blinking. Press CFM to
confirm the “100.0%” DO calibration.
If the reading is within the limits (±15% f.s.), the
meter stores the value (and adjusts the slope point).