Make sure the instrument has been calibrated before taking DO measurements.
DIRECT MEASUREMENT
To measure the DO of a sample using the Direct reading mode:
Press
and then to select DO measure mode.
Select the
Direct
reading mode (see
DO
Setup
).
Submerge the DO probe and tap
it repeatedly to remove any
air bubbles that may be trapped inside the sleeve. Allow time
for the reading to stabilize.
The measured DO value will be displayed together with the
temperature and pressure values.
Notes: For accurate DO measurements, a water movement of
0.3 m/s is required. This is to ensure that the oxygen-
depleted membrane surface is constantly replenished.
A moving stream will provide adequate circulation.
If the reading is out of range, “-----” will be displayed.
DIRECT/AUTOHOLD MEASUREMENT
To measure DO of a sample using the Direct/AutoHold reading mode:
Select the Direct/AutoHold reading mode (see
DO
Setup
).
If pressing
, the “AutoHold” indicator will start blink-
ing on the display until the stability criterion is reached. The DO
value will be frozen on the display, along with “AutoHold”
indicator.
To return to normal measure mode press
.