Instruction Manual HI 9210N Portable Waterproof pH Meter w w w . h a n n a i n s t .
Dear Customer, Thank you for choosing a Hanna Product. Please read this instruction manual carefully before using the instrument. If you require further technical information, do not hesitate to e-mail us at tech@hannainst.com. This instrument is in compliance with the directives. WARRANTY All Hanna Instruments meters are guaranteed for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions.
FUNCTIONAL DESCRIPTION SPECIFICATIONS 1 3 2 4 5 7 6 pH HI 9210N ATC pH METER OPERATIONAL GUIDE 8 9 RANGE pH: 0.00 - 14.00 °C: 0.0 -70.0 ON/OFF RANGE 1) 2) 3) 4) 5) 6) 7) 8) 9) Battery cover DIN connector for electrode Offset calibration trimmer Slope calibration trimmer Temperature calibration trimmer Mode indicator Liquid Crystal Display ON/OFF key, to turn the meter on or off RANGE key, to visualize the temperature 4 Range 0.00 to 14.00 pH / 0.0 to 70.0°C Resolution 0.01 pH / 0.
CALIBRATION For best accuracy, it is recommended to calibrate the instrument frequently. In addition a pH calibration should be performed: a) Whenever the pH electrode or the batteries are replaced b) At least once a month c) After testing aggressive chemicals d) When extreme accuracy is required pH • pH calibration is now complete. Replace the trimmer cover and tighten it to ensure a proper seal against water and humidity ingress. PREPARATION Pour small quantities of pH 7.01 and pH 4.01 (or pH 10.
pH VALUES AT VARIOUS TEMPERATURES Temperature has an effect on pH. The calibration buffer solutions are affected by temperature changes to a lesser degree than normal solutions. For pH calibration, please refer to the following chart. TEMP °C °F 0 32 5 41 10 50 15 59 68 20 77 25 86 30 95 35 40 104 45 113 50 122 55 131 60 140 65 149 70 158 4.01 4.01 4.00 4.00 4.00 4.00 4.01 4.02 4.03 4.04 4.05 4.06 4.07 4.09 4.11 4.12 pH VALUES 7.01 9.18 9.46 7.13 9.39 7.10 9.33 7.07 9.27 7.04 9.22 7.03 9.18 7.01 9.14 7.
IMPORTANT: After performing any of the cleaning procedures, rinse the electrode thoroughly with distilled water and soak it in HI 70300 storage solution for at least 1 hour before taking measurements. TROUBLESHOOTING Evaluate your electrode performance based on the following. • Noise (Readings fluctuate up and down) could be due to clogged/ dirty junction: Refer to the Cleaning Procedure above. • Dry membrane/junction: Soak in HI 70300 storage solution for at least one hour.
SALES AND TECHNICAL SERVICE CONTACTS Australia: Tel. (03) 9769.0666 • Fax (03) 9769.0699 China: Tel. (10) 88570068 • Fax (10) 88570060 Egypt: Tel. & Fax (02) 2758.683 Germany: Tel. (07851) 9129-0 • Fax (07851) 9129-99 Greece: Tel. (210) 823.5192 • Fax (210) 884.0210 Indonesia: Tel. (21) 4584.2941 • Fax (21) 4584.2942 Japan: Tel. (03) 3258.9565 • Fax (03) 3258.9567 Korea: Tel. (02) 2278.5147 • Fax (02) 2264.1729 Malaysia: Tel. (603) 5638.9940 • Fax (603) 5638.9829 Singapore: Tel. 6296.