User Guide

3130
Note To interrupt the calibration procedure press SETUP to restart
the procedure, or CAL to exit to normal operational mode.
Note If the process meter has never been calibrated or an EE-
PROM reset has occurred, the meter continues to perform
measurements. However, the user is informed of an EC or
TDS calibration requirement by a blinking “CAL” indication
(see “Startup” section).
Note The device must be calibrated within the temperature range
specified for the EC or TDS buffer solution.
CELL CONSTANT DIRECT SELECTION
Whenever the EC/TDS probe cell constant is known, it is
possible to directly calibrate the meter using that value.
Press CAL to enter calibration mode. The LCD will show
the default offset of 0.
Press LCD to display the current cell constant on the pri-
mary LCD (factory default value is 2.000 cm
-1
).
Press SETUP key.
Using !, " and # , enter the probe cell constant (the
value must be between 1.333 and 4.000 cm
-1
) and con-
firm by pressing CFM.
Note If the entered cell constant value is invalid, the “ERROR” indi-
cator blinks on the LCD.
Note Press SETUP before CFM to exit without changing the cell
constant.
CALIBRATION BUFFER DIRECT SELECTION
This feature allows to set a user-defined calibration point, in
order to perform calibration at a point different from the memo-
rized standards.
Press CAL to enter calibration mode. The LCD will show 0.
If the zero calibration cannot be performed,
"ERROR" will blink.
Cell constant calibration
Select the solution value on the
primary display by pressing !
or " if the selected range has
two possibilities (e.g. 5.000 and
12.880 mS).
Immerse the EC/TDS probe with the temperature sensor in
the selected solution. The level of solution must be higher
than the holes of the EC/TDS probe sleeve. Tap the EC/
TDS probe repeatedly on the bottom of the beaker and stir
to ensure that no air bubbles are trapped inside the sleeve.
When the reading is stable, "CAL"
will stop flashing (after about 30 sec-
onds) and the "READY" and "CFM"
indicators will blink.
Press CFM to confirm the calibra-
tion point; if the reading is close to
the selected solution, the meter
stores the reading.
If the reading is not close to the se-
lected solution, "ERROR" will blink.
Note A 2-point calibration is always suggested. However the EC/
TDS calibration can also be performed at 1 point. To cali-
brate offset only, just press CAL after confirmation (with CFM)
of the zero reading; the meter will return to normal opera-
tional mode. To have the cell constant calibrated first, press
the up or down arrow keys after entering the calibration pro-
cedure to skip to the next possible calibration buffer. In this
case, after confirmation of the cell constant, the meter will
ask for the offset calibration displaying zero on the LCD; press
CAL to exit or calibrate the offset, if desired.
Note The EC or TDS calibration value shown is referenced at 25°C
even if the reference temperature of 20°C has been selected.
Note During calibration, press LCD to display the cell constant
value on the primary display. Press LCD again to return to
calibration buffer visualization.