Instruction Manual EC214 EC215 - EC215R Bench Conductivity Meters http://www.hannainst.
Dear Customer, Thank you for choosing a Hanna Instruments Product. Please read this instruction manual carefully before using the instrument. This manual will provide you with all the necessary information for the correct use of the instrument, as well as a precise idea of its versatility in a wide range of applications. These instruments are in compliance with directives. TABLE OF CONTENTS PRELIMINARY EXAMINATION ............................................................
FUNCTIONAL DESCRIPTION EC 214 SPECIFICATIONS EC 214 Range Resolution Accuracy (@20°C / 68°F) Typical EMC Deviation Calibration Temperature Compensation Probe Power Supply Environment 1) 2) 3) 4) 5) 6) 7) Liquid Crystal Display (LCD) Conductivity calibration knob 8) 9) 10) 11) Temperature setting knob for manual compensation ON/OFF switch Power adapter socket Probe connector Dimensions Weight 0.0 to 199.9 µS/cm 0 to 1999 µS/cm 0.00 to 19.99 mS/cm 0.0 to 199.9 mS/cm 0.1 µS/cm 1 µS/cm 0.01 mS/cm 0.
FUNCTIONAL DESCRIPTION EC 215 SPECIFICATIONS EC 215 Range Resolution Accuracy (@20°C / 68°F) Typical EMC Deviation Calibration Temperature Compensation Probe (included) Analog Output (EC 215R only) 1) Liquid Crystal Display (LCD) 2) Conductivity calibration knob 3) 199.9 µS key, to select the range 0.0 to 199.9µS/cm 4) 19.99 mS key, to select the range 0.00 to 19.99 mS/cm 5) 199.9 mS key, to select the range 0.0 to 199.
OPERATIONAL GUIDE Power connection Plug the 12VDC adapter (HI710005 or HI710006) into the power supply socket (#10 on page 4; #9 on page 6). Note: make sure the main line is protected by a fuse. Probe connection Connect the conductivity probe to the socket provided (#11 on page 4; #10 on page 6). Note: Make sure that the instrument has been calibrated before taking conductivity measurements (see calibration paragraph). Note: If possible, use plastic beakers to minimize EMC interferences.
PROCEDURE FOR EC 214: 0 703 HI If you are measuring in the mS ranges, calibrate the meter using HI7030 (or HI8030) conductivity solution (12.88 mS @ 25 C) or HI7034 (or HI8034) conductivity solution (80 mS @ 25 C). For the µS ranges, use HI7031 (or HI8031) conductivity solution (1413 µS @25 C) when calibrating in the range from 0 to 1999 µS or HI7033 (or HI8033) conductivity solution (84 µS @ 25 C) when calibrating in the range from 0 to 199.9 µS.
• Select the appropriate range: "199.9 µS" for HI7033/HI8033, "1999µS" for HI7031/HI8031, "19.99 mS" for HI7030/HI8030, "199.9 mS" for HI7034/HI8034. Note: If the display shows "1", there is an over-range condition. Select the next higher range. • Allow a few minutes for the reading to stabilize and adjust the “CALIBRATION” knob to read (on the LCD) the value of the buffer solution at 25 C (77 F),e.g. 12.88 mS/cm.
DETERMINATION OF THE TEMPERATURE COEFFICIENT OF A SOLUTION (EC 215 & EC 215R) • Immerse the probe into a sample of the solution and adjust the “TEMPERATURE COEFFICIENT” knob to 0% (i.e. no compensation). • Condition the sample and probe at 25 C and note the conductivity reading C25. • Condition the sample and probe to a temperature t C which is approximately 5 C to 10 C different from 25 C and note the conductivity reading Ct.
ACCESSORIES CONDUCTIVITY BUFFER SOLUTIONS: HI 7030L HI 7030M HI 7031L HI 7031M HI 7033L HI 7033M HI 7034L HI 7034M HI 7035L HI 7035M HI 7039L HI 7039M 12880 µS/cm (µmho/cm), 460mL 12880 µS/cm (µmho/cm), 230mL 1413 µS/cm (µmho/cm), 460mL 1413 µS/cm (µmho/cm), 230mL 84 µS/cm (µmho/cm), 460 mL 84 µS/cm (µmho/cm), 230 mL 80000 µS/cm (µmho/cm), 460mL 80000 µS/cm (µmho/cm), 230mL 111800 µS/cm (µmho/cm), 460mL 111800 µS/cm (µmho/cm), 230mL 5000 µS/cm (µmho/cm), 460mL 5000 µS/cm (µmho/cm), 230mL OTHER ACCESSORIES
WARRANTY CE DECLARATION OF CONFORMITY All Hanna Instruments meters are guaranteed for two years against defects in workmanship and materials when used for their intended purpose and maintained according to instructions. The electrodes and the probes are guaranteed for a period of six months. This warranty is limited to repair or replacement free of charge. Damage due to accident, misuse, tampering or lack of prescribed maintenance are not covered.