Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
Fill one graduated mixing cylinder up to the 25
mL mark with the sample.
Add the content of one
packet of HI 93730A mo-
lybdenum reagent, close
the cylinder and invert it
several times until dissolu-
tion is complete.
Add the content of one
packet of HI 93730B mo-
lybdenum reagent to the
cylinder, close and invert
it several times until dis-
solution is complete.
Add the content of one
packet of HI 93730C mo-
lybdenum reagent to the
cylinder, close and shake
it vigorously.
Fill an empty cuvet with 10 mL of
sample up to the mark and re-
place the cap.
Insert the cuvet into the instrument.
25 mL
Molybdenum
10 mL