User Guide

92
Manganese LR
Add 0.2 mL of the
HI 93748B Alkaline-cya-
nide reagent solution to
each cuvet, replace the
caps and shake gently.
Add 1 mL of the
HI 93748C 0.1% PAN in-
dicator solution to each
cuvet, replace the caps
and shake gently.
Place the cuvet with the reacted deion-
ized water (blank) into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press TIMER and the display will show the countdown prior to
zeroing the blank. Alternatively wait for 2 minutes and then
press ZERO. "SIP" will blink during zeroing.
Wait for a few seconds and the display will
show "-0.0-". Now the meter is zeroed and
ready for measurement.
Insert the second cuvet with the reacted
sample into the instrument.