User Guide

74Total Hardness MR
Add 2 drops of HI 93735B
Buffer reagent
Replace the cap and shake gently to mix.
Wait for 15 seconds.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and "SIP" will blink on the display.
Wait for a few seconds and the display will
show "-0.0-". Now the meter is zeroed and
ready for measurement.
Remove the cuvet and add the content of 1 packet
of HI 93735C Fixing reagent.
Replace the cap and shake
gently till dissolution is com-
plete.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
x 2