TOTAL HARDNESS HIGH RANGE TOTAL HARDNESS HIGH RANGE
TOTAL HARDNESS HIGH RANGE TOTAL HARDNESS HIGH RANGE
TOTAL HARDNESS HIGH RANGE
SPECIFICATIONS
Range 400 to 750 mg/L
Resolution 5 mg/L
Accuracy ±10 mg/L ±2% of reading
Typical EMC ±5 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 466 nm
Method Adaptation of the EPA method 130.1. The reac-
tion between calcium, magnesium and reagents
causes a red-violet tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93735A-HR Hardness Indicator reagent HR 9.5 mL
HI 93735B-0 Hardness Buffer reagent 2 drops
HI 93735C-0 Fixing reagent 1 packet
REAGENT SETS
HI 93735-02 Reagents for 100 tests High Range
(400 to 750 mg/L)
HI 93735-0 Reagents for 100 tests (0 to 750 mg/L)
For other accessories see page 141.
MEASUREMENT PROCEDURE
Select the program number corresponding to Total Hardness HR
on the secondary LCD by pressing PROGRAM W and V.
Using the syringe, add exactly 0.5 mL of
unreacted sample to the cuvet.
With the plastic pipette fill the cuvet
up to the 10 mL mark adding HI
93735A-HR Indicator reagent.
Total Hardness HR