12
3
6
9
10"
Remove the cuvet and add 1
level spoon of HI 93714A Cya-
nide Reagent. Remember to close
the reagent bottle immediately
after use.
Note: Pay attention to the way the spoon is filled:
- do not press the powder;
- do not overfill it.
Place the HDPE plastic stopper
and cap immediately, to prevent
the escape of chlorine gas which
is developed during the reaction,
and shake gently for 30 seconds.
Wait for 30 seconds leaving the
cuvet tightly capped and undis-
turbed, then add the content of
one packet of HI 93714B re-
agent and shake gently for 10
seconds.
Immediately add the content of
one packet of HI 93714C re-
agent, replace the cap and shake
vigorously for 20 seconds.
Reinsert the cuvet into the instrument.