COPPER LOW RANGECOPPER LOW RANGE
COPPER LOW RANGECOPPER LOW RANGE
COPPER LOW RANGE
SPECIFICATIONS
Range 0 to 1000 µg/L
Resolution 1 µg/L
Accuracy ±10 µg/L ±5% of reading
Typical EMC ±1 µg/L
Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 575 nm
Method Adaptation of the EPA method. The reaction
between copper and the bicinchoninate reagent
causes a purple tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93747-0 Bicinchoninate 1 packet
REAGENT SETS
HI 93747-01 Reagents for 100 tests
HI 93747-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
Select the program number corresponding to Copper LR on the
secondary LCD by pressing PROGRAM W and V.
Fill the cuvet with 10 mL of unreacted sample (up
to the mark) and replace the cap.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and “SIP” will blink on the display.
Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed and
ready for measurement.
Copper LR
10 mL