User Guide

47
Chromium VI HR
Remove the cuvet and add the
content of one packet of HI
93723 reagent. Replace the cap
and shake vigorously for about
10 seconds.
Reinsert the cuvet into the instrument.
Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 6 minutes and press READ
DIRECT. In both cases "SIP" will blink during measurement.
The instrument directly displays concentration in µg/L of chromium
VI on the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by:
Vanadium above 1 ppm. However, waiting 10 minutes before
reading, the interference is removed
Iron above 1 ppm
Mercurous and mercuric ions cause slight inhibition of the reaction.