SPECIFICATIONS
Range 0.00 to 8.00 mg/L
Resolution 0.01 mg/L
Accuracy ±0.08 mg/L ±3% of reading
Typical EMC ±0.01 mg/L
Deviation
Light Source Tungsten lamp with narrow band interference
filter @ 525 nm
Method Adaptation of the Standard Methods for the
Examination of Water and Wastewater, 18
th
edition, DPD method. The reaction between bro-
mine and the reagent causes a pink tint in the
sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93716-0 DPD Reagent 1 packet
REAGENT SETS
HI 93716-01 Reagents for 100 tests
HI 93716-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
Select the program number corresponding to Bromine on the
secondary LCD by pressing PROGRAM W and V.
Fill the cuvet with 10 mL of unreacted sample (up
to the mark) and replace the cap.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and “SIP” will blink on the display.
10 mL