User Guide

22
Aluminum
Fill two cuvets with 10 mL of sample
each (up to the mark).
Add the content of one HI 93712C
Bleaching powder packet to one of
the two cuvets. Replace the cap
and shake vigorously until dissolution
is complete. This is the blank.
Place the blank into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press TIMER and the display will show the countdown prior to
zeroing the blank. Alternatively wait for 15 minutes and then
press ZERO. "SIP" will blink during zeroing.
Wait for a few seconds and the display will
show "-0.0-". Now the meter is zeroed and
ready for measurement.
Remove the blank and insert the other
cuvet into the instrument.
Press READ DIRECT.
"SIP" will blink during
measurement.
The instrument directly displays concentration in mg/L of aluminum
on the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by:
Iron above 20 mg/L Alkalinity above 1000 mg/L
Phosphate above 50 mg/L Fluoride must be absent
10 mL10 mL
# 1
# 1
# 2
# 1
# 2