135 Zinc
Fill one cuvet with 10 mL of the reacted
sample up to the mark.
Place the cap and insert the cuvet into
the cell and ensure that the notch on
the cap is positioned securely into the
groove.
Press ZERO and “SIP” will blink on the display.
Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
Remove the cuvet and add 0.5 mL of HI 93731B
cyclohexanone to the cuvet.
Note: To prevent any contamination from
the polycarbonate cap, prior to replacing it,
close the sample cuvet with the supplied
HDPE plastic stopper.
Replace the cap and mix the sample for
15 seconds.
Insert the sample into the instru-
ment.
10 mL