User Guide

129
Wait for 1 minute. This is the blank.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and “SIP” will blink on the display.
Wait for a few seconds and the display
will show “-0.0-”. Now the meter is
zeroed and ready for measurement.
Remove the cuvet and add the
content of one packet of HI 93705C
Amino acid reagent and shake until
it is completely dissolved.
Reinsert the cuvet into the instrument.
Press TIMER and the display will show the countdown prior to the
measurement. Alternatively, wait for exactly 3 minutes and press
READ DIRECT. In both cases “SIP” will blink during measurement.
The instrument directly displays concentration in mg/L of silica
(SiO
2
) on the Liquid Crystal Display.
12
3
6
9
1’
Silica