Phosphorus
x 10
Wait for a few seconds and the display
will show "-0.0-". Now the meter is
zeroed and ready for measurement.
Remove the cuvet.
Add 10 drops of HI 93706A Molybdate
reagent.
Add the content of one packet of
HI 93706B-0 Phosphorus Reagent B
(Amino Acid) to the cuvet. Replace
the cap and shake gently until
dissolution is complete.
Reinsert the cuvet into the instrument.
Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 5 minutes and press READ
DIRECT. In both cases "SIP" will blink during measurement.
The instrument directly displays concentration in mg/L of phosphorus
on the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by:
Sulfide
Chloride above 150000 mg/L
Calcium above 10000 mg/L as CaCO
3
Magnesium above 40000 mg/L as CaCO
3
Ferrous iron above 100 mg/L