User Guide

126
PHOSPHORUSPHOSPHORUS
PHOSPHORUSPHOSPHORUS
PHOSPHORUS
Phosphorus
SPECIFICATIONS
Range 0.0 to 15.0 mg/L
Resolution 0.1 mg/L
Accuracy ±0.3 mg/L ±4% of reading
Typical EMC dev. ±0.2 mg/L
Light Source Tungsten lamp with narrow band interference
filter @ 525 nm
Method Adaptation of the Standard Methods for the
Examination of Water and Wastewater, 18
th
edition, Amino Acid method. The reaction be-
tween phosphorus and the reagents causes a blue
tint in the sample.
REQUIRED REAGENTS
Code Description Quantity
HI 93706A-0 Molybdate 10 drops
HI 93706B-0 Amino Acid Powder 1 packet
REAGENT SETS
HI 93706-01 Reagents for 100 tests
HI 93706-03 Reagents for 300 tests
For other accessories see page 141.
MEASUREMENT PROCEDURE
Select the program number corresponding to Phosphorus on the
secondary LCD by pressing PROGRAM W and V.
Fill the cuvet with 10 mL of unreacted
sample (up to the mark) and replace
the cap.
Place the cuvet into the holder and
ensure that the notch on the cap is
positioned securely into the groove.
Press ZERO and “SIP” will blink on the display.
10 mL