125 Phosphate LR
Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
Remove the cuvet and add the content
of one packet of HI 93713 reagent.
Replace the cap and shake gently (for
about 2 minutes) until the powder is
completely dissolved.
Reinsert the cuvet into the instrument.
Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 3 minutes and press READ
DIRECT. In both cases "SIP" will blink during measurement.
The instrument directly displays concentration in mg/L of phosphate
on the Liquid Crystal Display.
To convert the reading to mg/L of P
2
O
5
, multiply by a factor of
0.747.
To convert the reading to mg/L of phosphorus (P) concentration,
multiply by a factor of 0.326.
INTERFERENCES
Interference may be caused by:
Iron above 50 mg/L
Silica above 50 mg/L
Silicate above 10 mg/L
Copper above 10 mg/L
Hydrogen sulfide, arsenate, turbid sample and highly buffered
samples also interfere.