User Guide

123
Phosphate HR
Press ZERO and “SIP” will blink on the display.
Wait for a few seconds and the display will
show "-0.0-". Now the meter is zeroed and
ready for measurement.
Remove the cuvet.
Add 10 drops of HI 93717A
Molybdate reagent.
Add to the cuvet 0.5 mL of the
solution prepared in the HI
93717B bottle. Replace the cap
and mix the solution.
Reinsert the cuvet into the instrument.
Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 5 minutes and press READ
DIRECT. In both cases "SIP" will blink during measurement.
The instrument directly displays concentration in mg/L of phos-
phate (PO
4
3–
) on the Liquid Crystal Display.
INTERFERENCES
Interference may be caused by:
Sulfide
Chloride above 150000 mg/L
Calcium above 10000 mg/L as CaCO
3
Magnesium above 40000 mg/L as CaCO
3
Ferrous iron above 100 mg/L
x 10