User Guide

119
Wait for a few seconds and the display will
show “-0.0-”. Now the meter is zeroed
and ready for measurement.
Remove the cuvet.
Fill another cuvet up to the mark with 10
mL of the reacted sample and replace the
cap.
Reinsert the cuvet into the instrument.
Press READ DIRECT and "SIP" will blink during measurement.
The instrument will then directly display the concentration of dis-
solved oxygen in mg/L on the Liquid Crystal Display.
INTERFERENCES
Interferences may be caused by reducing and oxidizing materials.
Dissolved Oxygen