User Guide

107
Remove the cuvet.
Add the content of one packet of
HI 93707 reagent. Replace the
cap and shake gently for about
15 seconds.
Reinsert the cuvet into the instrument.
Press TIMER and the display will show the countdown prior to the
measurement or, alternatively, wait for 6 minutes and press READ
DIRECT. In both cases “SIP” will blink during measurement.
The instrument directly displays concentration in mg/L of nitrite
nitrogen.
To convert the NO
2
-N concentration to the nitrite ion concentra-
tion (NO
2
), multiply the reading by factor of 3.29.
To convert the NO
2
-N concentration to sodium nitrite concentra-
tion (NaNO
2
), multiply the reading by factor of 4.93.
INTERFERENCES
Interference may be caused by the following ions:
ferrous, ferric, cupric, mercurous, silver, antimonious, bismuth, auric,
lead, metavanadate and chloroplatinate.
Strongly reducing and oxidizing reagents.
High levels of nitrate (above 100 mg/L) could yield falsely high
readings due to a minute amount of reduction to nitrite that could
occur at these levels.