Instruction Manual C 99 & C 200 Series Multiparameter Bench Photometers Manufacturers since 1978 1 These Instruments are in Compliance with the CE Directives
Dear Customer, Thank you for choosing a Hanna product. Please read this instruction manual carefully before using the meter. This manual will provide you with the necessary information for the correct use of the instrument. If you need additional technical information, do not hesitate to e-mail us at tech@hannainst.com. These instruments are in compliance with directives. TABLE OF CONTENTS TOTAL HARDNESS LOW RANGE ......... 76 HYDRAZINE ................................... 79 IODINE ........................
PRELIMINARY EXAMINATION Remove the instrument from the packing material and examine it carefully to make sure that no damage has occurred during shipment. If there is any damage, notify your Dealer. Each Meter is supplied complete with: • Two Sample Cuvets and Caps* • One Transport Cap • Two 9 V Batteries • 12 VDC Transformer (HI 710005 or HI 710006) Note: Save all packing material until you are sure that the instrument functions correctly.
SPECIFICATIONS Life of the instrument Silicon Photocell 0 to 50°C (32 to 122°F); max 95% RH non-condensing Power Supply 2 x 9 V batteries / 12 to 20 VDC through voltage adapter Auto-Shut off After 10' of non-use Dimensions 230 x 165 x 70 mm (9.0 x 6.5 x 2.8") Weight 640 g (22.6 oz.) For specifications related to each single parameter (e.g. range, accuracy, etc.), refer to the related measurement section.
Given that the absorption of a compound strictly depends on the wavelength of the incident light beam, a narrow spectral bandwidth should be selected as well as a proper central wavelength to optimize measurements. The optical system of Hanna's C 99 & C 200 multiparameter photometers is based on special subminiature tungsten lamps and narrow-band interference filters to guarantee both high performance and reliable results. Four measuring channels (at four different wavelengths) allow a wide range of tests.
FUNCTIONAL DESCRIPTION FRONT PANEL 1) 2) 3) 4) 5) Cuvet Holder Dual Level Liquid Crystal Display Programs List READ DIRECT, to perform measurement immediately TIMER, to perform measurements after a preprogrammed countdown 6) ZERO, to zero the meter prior to measurement 7) Program W and V, to select the desired parameter 8) ON/OFF, to turn the meter on and off REAR PANEL 1) Power Supply 12 VDC 2.
GUIDE TO DISPLAY CODES Note: The secondary LCD below shows a generic "P– –", whereas the meter will indicate the exact program number (e.g. in C 200, "P1" for Aluminum). This indicates that the meter is in a ready state and zeroing can be performed. Sampling in progress. This flashing prompt appears each time the meter is performing a measurement. The microprocessor is adjusting the light level, indicated by a scrolling "SIP".
Light over range. The cuvet is not inserted correctly and an eccess ambient light is reaching the detector. If the cuvet is properly inserted, then contact your dealer or the nearest Hanna Customer Service Center. The lamp is not working properly. Contact your dealer or the nearest Hanna Customer Service Center. The lamp is not working properly. Contact your dealer or the nearest Hanna Customer Service Center. This indicates that the meter has lost its configuration.
b) on sample reading: There is too much light for the sample measurement. Please check if the right sample cuvet is inserted. The sample and zero cuvet are inverted. A zero reading was not taken. Follow the instruction described in the measurement procedures for zeroing the meter. Under range. A blinking "0.00" indicates that the sample absorbs less light than the zero reference. Check the procedure and make sure that you use the same cuvet for reference (zero) and measurement.
TIPS FOR AN ACCURATE MEASUREMENT The instructions listed below should be carefully followed during testing to ensure best accuracy. • Color or suspended matter in large amounts may cause interference, therefore, these should be removed by treatment with active carbon and by prior filtration. • For a correct filling of the cuvet: the liquid in the cuvet forms a convexity on the top; the bottom of this convex- 10 mL ity must be at the same level of the 10 mL mark. • In order to measure exactly e.g. 0.
• Proper use of the powder reagent packet: (a) use scissors to open the powder packet; (b) push the edges of the packet to form a spout; (c) pour out the content of the packet. (a) (b) (c) • It is important that the sample does not contain any debris. This would corrupt the readings. • Do not let the reacted sample stand too long after reagent is added, or accuracy will be lost.
PARAMETERS REFERENCE TABLES C99 - LABORATORIES, with COD Code Parameter Page 1 Aluminum 2 Ammonia MR 21 3 Ammonia LR 4 Bromine 27 29 5 Free Chlorine 6 Total Chlorine 7 Chlorine Dioxide 31 37 43 8 Chromium VI HR 9 Chromium VI LR 46 48 Code Parameter 22 Manganese HR 23 Manganese LR 25 10 Color of Water 11 Copper HR 12 Copper LR 50 52 54 13 Cyanide 14 Cyanuric Acid 56 59 15 Fluoride 16 Ca Hardness 17 Mg Hardness 61 64 67 18 Hydrazine 79 19 Iodine 82 20 Iron HR 21 Iron LR 84 86 Page 89 9
PARAMETERS REFERENCE TABLES C 200 - LABORATORIES Code Parameter Page 1 Aluminum 21 2 Ammonia MR 3 Ammonia LR 25 27 4 Bromine 5 Free Chlorine 6 Total Chlorine 29 31 37 7 Chlorine Dioxide 8 Chromium VI HR 43 46 9 Chromium VI LR 10 Color of Water 11 Copper HR 48 50 52 12 Copper LR 13 Cyanide 54 56 14 Cyanuric Acid 15 Fluoride 16 Ca Hardness 59 61 64 17 Mg Hardness 18 Hydrazine 67 79 Code Parameter Page 19 Iodine 20 Iron HR 21 Iron LR 82 84 86 22 Manganese HR 23 Manganese LR 89 91 24 Mo
C 205 - BOILERS & COOLING TOWERS Code Parameter Page Code Parameter Page 1 Aluminum 21 13 Iron HR 84 2 Ammonia MR 3 Ammonia LR 25 27 14 Iron LR 15 Molybdenum 86 94 4 Bromine 5 Free Chlorine 6 Total Chlorine 29 31 37 16 Nitrate 17 Nitrite HR 18 Nitrite LR 7 Chlorine Dioxide 8 Chromium VI HR 43 46 19 Oxygen, Dissolved 117 20 pH 120 9 Chromium VI LR 10 Copper HR 48 52 21 Phosphate HR 22 Phosphate LR 122 124 11 Copper LR 12 Hydrazine 54 79 23 Silica 24 Zinc 128 134 102 104 106 C 206 -
C 207 - INDUSTRIAL WASTEWATER Code Parameter Page Code Parameter Page 1 Aluminum 21 11 Nickel HR 97 2 Free Chlorine 3 Total Chlorine 31 37 12 Nickel LR 13 Nitrate 99 102 4 Color of Water 5 Copper HR 6 Copper LR 50 52 54 14 Oxygen, Dissolved 117 15 pH 120 16 Phosphate HR 122 7 Fluoride 8 Manganese HR 61 89 17 Phosphate LR 18 Phosphorus 124 126 9 Manganese LR 10 Molybdenum 91 94 19 Silver 20 Zinc 131 134 C 208 - WATER CONDITIONING Code Parameter Page Code Parameter Page 1 Ammonia MR
C 209 - EDUCATION Code Parameter Page Code Parameter Page 1 Ammonia MR 2 Ammonia LR 3 Free Chlorine 25 27 31 11 Nitrite HR 104 4 Total Chlorine 5 Chromium VI HR 37 46 6 Chromium VI LR 7 Color of Water 48 50 8 Copper HR 9 Copper LR 10 Nitrate 12 Nitrite LR 106 13 Oxygen, Dissolved 117 14 pH 120 52 54 102 15 Phosphate HR 16 Phosphate LR 122 124 17 Phosphorus 18 Silica 126 128 19 Silver 20 Zinc 131 134 C 210 - PULP & PAPER MILLS Code Parameter 1 Aluminum 2 Free Chlorine 3 Total Chlorine 4
C 211 - CHEMICAL MANUFACTURERS Code Parameter 1 Aluminum Page 21 Code Parameter 12 Molybdenum Page 94 2 Ammonia MR 3 Ammonia LR 4 Chromium VI HR 25 27 46 13 Nickel HR 14 Nickel LR 15 pH 97 99 120 5 Chromium VI LR 6 Copper HR 48 52 16 Phosphate HR 17 Phosphate LR 122 124 7 Copper LR 8 Cyanuric Acid 9 Iodine 54 59 82 18 Phosphorus 19 Silica 20 Silver 126 128 131 84 86 21 Zinc 134 10 Iron HR 11 Iron LR C 212 - POWER PLANT UTILITIES Code Parameter Page Code Parameter Page 1 Ammonia MR 2
C 213 - MUNICIPAL WASTEWATER Code Parameter Page 1 Aluminum 2 Ammonia MR 3 Ammonia LR 25 27 4 Bromine 5 Free Chlorine 29 31 6 Total Chlorine 7 Chromium VI HR 8 Chromium VI LR 37 46 48 Code Parameter 21 9 Color of Water 10 Copper HR 50 52 11 Copper LR 12 Iodine 54 82 Page 13 Nickel HR 14 Nickel LR 97 99 15 Nitrate 16 Nitrite HR 17 Nitrite LR 102 104 106 18 Oxygen, Dissolved 117 19 pH 120 20 Phosphate HR 21 Phosphate LR 122 124 22 Phosphorus 23 Silver 24 Zinc 126 131 134 C 216 - SWIMMIN
OPERATIONAL GUIDE POWER CONNECTION Plug the 12VDC adapter (HI 710005 - 110VDC, or HI 710006 220VDC) into the DC socket. Plug the adapter into the outlet. Alternatively, remove the battery cover on the back of the meter; attach 2 fresh 9V batteries and replace the cover. Note: Insure the main line is surge protected. Note: Always turn the meter off before unplugging it to insure no data is lost. MEASUREMENT PROCEDURE • Turn the meter on by pressing ON/OFF.
• After the desired program number appears on the secondary display, follow the measurement procedure described in the related chapter. • Select a new parameter measurement procedure by pressing the PROGRAM W and PROGRAM V keys. Note: in the following measurement sections, a generic "P– –" will be placed on the secondary LCD instead of the exact related message (e.g. in C 200, "P1" for Aluminum). • Before performing a test read carefully all the instructions related to the selected parameter.
ALUMINUM SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 1.00 mg/L 0.01 mg/L ±0.02 mg/L ±4% of reading ±0.01 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the aluminon method. The reaction between aluminum and reagents causes a reddish tint in the sample.
• Fill two cuvets with 10 mL of sample each (up to the mark). • Add the content of one HI 93712C Bleaching powder packet to one of the two cuvets. Replace the cap and shake vigorously until dissolution is complete. This is the blank. • Place the blank into the holder and ensure that the notch on the cap is positioned securely into the groove. 10 mL #1 10 mL #2 #1 #1 • Press TIMER and the display will show the countdown prior to zeroing the blank.
AMMONIA HIGH RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.0 to 50.0 mg/L 0.1 mg/L ±0.5 mg/L ±5% of reading ±0.1 mg/L Tungsten lamp with narrow band interference filter @ 420 nm Adaptation of the ASTM Manual of Water and Environmental Technology, D1426-92, Nessler method. The reaction between ammonia and reagents causes a yellow tint in the sample.
• Press ZERO and “SIP” will blink on the display. • Wait for a few seconds and the display will show “-0.0-”. Now the meter is zeroed and ready for measurement. • Remove the cuvet. • Add 4 drops of HI 93733A Nessler Reagent. Replace the cap and mix the solution. x4 • Reinsert the cuvet into the instrument. • Press TIMER and the display will show the countdown prior to measurement or, alternatively, wait for 3 minutes and 30 seconds and then press READ DIRECT.
AMMONIA MEDIUM RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 10.00 mg/L 0.01 mg/L ±0.05 mg/L ±5% of reading ±0.01 mg/L Tungsten lamp with narrow band interference filter @ 420 nm Adaptation of the ASTM Manual of Water and Environmental Technology, D1426-92, Nessler method. The reaction between ammonia and reagents causes a yellow tint in the sample.
• Remove the cuvet. • Add 4 drops of the First reagent (6 drops in case of seawater analysis). Replace the cap and mix the solution. • Add 4 drops of the Second reagent (10 drops in case of seawater analysis). Replace the cap and mix the solution. x4 x4 • Reinsert the cuvet into the instrument. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 3 minutes and 30 seconds and press READ DIRECT. In both cases “SIP” will blink during measurement.
AMMONIA LOW RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 3.00 mg/L 0.01 mg/L ±0.04 mg/L ±4% of reading ±0.01 mg/L Tungsten lamp with narrow band interference filter @ 420 nm Adaptation of the ASTM Manual of Water and Environmental Technology, D1426-92, Nessler method. The reaction between ammonia and reagents causes a yellow tint in the sample.
• Remove the cuvet. • Add 4 drops of the First reagent (6 drops in case of seawater analysis). Replace the cap and mix the solution. • Add 4 drops of the Second reagent (10 drops in case of seawater analysis). Replace the cap and mix the solution. x4 x4 • Reinsert the cuvet into the instrument. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 3 minutes and 30 seconds and press READ DIRECT. In both cases “SIP” will blink during measurement.
BROMINE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 8.00 mg/L 0.01 mg/L ±0.08 mg/L ±3% of reading ±0.01 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, DPD method. The reaction between bromine and the reagent causes a pink tint in the sample.
• Wait for a few seconds and the display will show “-0.0-”. Now the meter is zeroed and ready for measurement. • Remove the cuvet and add the content of one packet of HI 93716 reagent. Replace the cap and shake gently for about 20 seconds to dissolve most of the reagent. • Reinsert the cuvet into the instrument. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 2 minutes and 30 seconds and press READ DIRECT.
FREE CHLORINE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 2.50 mg/L 0.01 mg/L ±0.03 mg/L ±3% of reading ±0.01 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the EPA DPD method 330.5. The reaction between free chlorine and the DPD reagent causes a pink tint in the sample.
• Press ZERO and “SIP” will blink on the display. • Wait for a few seconds and the display will show “-0.0-”. Now the meter is zeroed and ready for measurement. • Remove the cuvet. Powder reagents procedure • Add the content of one packet of HI 93701 DPD reagent. Replace the cap and shake gently for 20 seconds (or 2 minutes in case of seawater analysis). • Wait for a minute to allow the undissolved reagent to precipitate and reinsert the cuvet into the instrument.
Liquid reagents procedure • To an empty cuvet add 3 drops of HI 93701A-F DPD1 indicator and 3 drops of HI 93701B-F DPD1 buffer. Swirl gently to mix, and immediately add 10 mL of unreacted sample. Replace the cap and shake gently again. x3 x3 • Reinsert the cuvet into the instrument. • Press READ DIRECT and the display will show “SIP” during measurement. • The instrument directly displays concentration in mg/L of free chlorine on the Liquid Crystal Display.
FREE CHLORINE HIGH RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.0 to 10.0 mg/L 0.1 mg/L ±0.1 mg/L ±3% of reading ±0.1 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the EPA DPD method 330.5. The reaction between free chlorine and the DPD reagent causes a pink tint in the sample.
Note: To measure exactly 5 mL of reagent with the syringe, push the plunger completely into the syringe and insert the tip into HI 93734B reagent bottle. Pull the plunger out until the lower edge of the seal is on the 5 mL mark of the syringe. probable level of liquid taken up by syringe • Fill the cuvet up to the 10 mL mark with 5 mL of unreacted sample, using the 3 mL plastic pipette. Note: rinse the 3 mL plastic pipette 2 or 3 times with sample before adding it to the cuvet with reagent.
• Remove the cuvet. • Add the content of one packet of HI 93701 DPD reagent. Replace the cap and shake gently for 20 seconds (or 2 minutes in case of seawater analysis). • Wait for a minute to allow the undissolved reagent to precipitate and reinsert the cuvet into the instrument. • Press READ DIRECT and the display will show “SIP” during measurement. • The instrument directly displays concentration in mg/L of free chlorine on the Liquid Crystal Display.
TOTAL CHLORINE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 3.50 mg/L 0.01 mg/L ±0.03 mg/L ±3% of reading ±0.01 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the EPA DPD method 330.5. The reaction between the chlorine and the DPD reagent causes a pink tint in the sample.
• Press ZERO and “SIP” will blink on the display. • Wait for a few seconds and the display will show “-0.0-”. Now the meter is zeroed and ready for measurement. • Remove the cuvet. Powder reagents procedure • Add 1 packet of HI 93711 reagent. Replace the cap and shake gently for 20 seconds (or 2 minutes in case of seawater analysis). • Reinsert the cuvet into the instrument.
mL of unreacted sample. Replace the cap and shake gently again. • Reinsert the cuvet into the instrument. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 2 minutes and 30 seconds and press READ DIRECT. In both cases “SIP” will blink during measurement. • The instrument directly displays concentration in mg/L of total chlorine on the Liquid Crystal Display.
TOTAL CHLORINE HIGH RANGE SPECIFICATIONS 0.0 to 10.0 mg/L 0.1 mg/L ±0.1 mg/L ±3% of reading ±0.1 mg/L Range Resolution Accuracy Typical EMC Deviation Light Source Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the EPA DPD method 330.5. The reaction between free chlorine and the DPD reagent causes a pink tint in the sample.
Note: To measure exactly 5 mL of reagent with the syringe, push the plunger completely into the syringe and insert the tip into HI 93734B reagent bottle. Pull the plunger out until the lower edge of the seal is on the 5 mL mark of the syringe. probable level of liquid taken up by syringe • Fill the cuvet up to the 10 mL mark with 5 mL of unreacted sample, using the 3 mL plastic pipette. Note: rinse the 3 mL plastic pipette 2 or 3 times with sample before adding it to the cuvet with reagent.
• Remove the cuvet. • Add 3 drops of HI 93734C reagent to the cuvet. • Add the content of one packet of HI 93701 DPD reagent to the cuvet. Replace the cap and shake gently for 20 seconds (or 2 minutes in case of seawater analysis). x3 • Reinsert the cuvet into the instrument. • Press TIMER and the display will show a countdown prior to the measurement or, alternatively, wait for 2 minutes and 30 seconds and press READ DIRECT. The display will show “SIP” during measurement.
CHLORINE DIOXIDE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 2.00 mg/L 0.01 mg/L ±0.10 mg/L ±5% of reading ±0.01 mg/L Tungsten lamp with narrow band interference filter @ 575 nm Adaptation of the Chlorophenol Red method. The reaction between chlorine dioxide and reagents causes a colorless to purple tint in the sample.
• Add the content of one packet of HI 93738B dechlorinating reagent to only one of the two cylinders (#1), close and invert it several times until it is totally dissolved. This is the blank. #1 • Add precisely 0.5 mL of HI 93738C chlorine dioxide reagent to each cylinder (#1 & #2), close them and invert several times to mix. #1 & #2 #1 #2 • Add 0.5 mL of HI 93738D chlorine dioxide reagent to each cylinder (#1 & #2), close them and invert several times to mix. Cylinder #2 is the reacted sample.
• Press ZERO and "SIP" will blink on the display. • Wait for a few seconds and the display will show "-0.0-". Now the meter is zeroed and ready for measurement. • Fill another cuvet with 10 mL of the reacted sample (#2) up to the mark and replace the cap. • Insert the sample into the instrument. #2 10 mL #2 • Press READ DIRECT and "SIP" will blink during measurement. • The instrument directly displays concentration in mg/L of chlorine dioxide on the Liquid Crystal Display.
CHROMIUM VI HIGH RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0 to 1000 µg/L 1 µg/L ±5 µg/L ±4% of reading ±1 µg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the ASTM Manual of Water and Environmental Technology, D1687-92, Diphenylcarbohydrazide method. The reaction between chromium VI and the reagent causes a purple tint in the sample.
• Remove the cuvet and add the content of one packet of HI 93723 reagent. Replace the cap and shake vigorously for about 10 seconds. • Reinsert the cuvet into the instrument. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 6 minutes and press READ DIRECT. In both cases "SIP" will blink during measurement. • The instrument directly displays concentration in µg/L of chromium VI on the Liquid Crystal Display.
CHROMIUM VI LOW RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0 to 300 µg/L 1 µg/L ±1 µg/L ±4% of reading ±1 µg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the ASTM Manual of Water and Environmental Technology, D1687-92, Diphenylcarbohydrazide method. The reaction between chromium VI and the reagent causes a purple tint in the sample.
• Remove the cuvet and add the content of one packet of HI 93749 reagent. Replace the cap and shake vigorously for about 10 seconds. • Reinsert the cuvet into the instrument. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 6 minutes and press READ DIRECT. In both cases "SIP" will blink during measurement. • The instrument directly displays concentration in µg/L of Chromium VI on the Liquid Crystal Display.
COLOR OF WATER SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0 to 500 PCU (Platinum Cobalt Units) 1 PCU ±10 PCU ±5% of reading ± 1 PCU Tungsten lamp with narrow band interference filter @ 420 nm Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Colorimetric Platinum Cobalt method. REQUIRED ACCESSORIES 0.45 µm membrane for true color measurement. For other accessories see page 141.
• Fill another cuvet up to the mark with unfiltered sample and replace the cap. This is the apparent color. • Filter 10 mL of sample through a filter with a 0.45 µm membrane into the third cuvet, up to the 10 mL mark and replace the cap. This is the true color. 10 mL #2 #3 • Insert the apparent color cuvet (# 2) into the instrument and ensure that the notch on the cap is positioned securely into the groove. #2 • Press READ DIRECT and "SIP" will blink on the display.
COPPER HIGH RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 5.00 mg/L 0.01 mg/L ±0.02 mg/L ±4% of reading ±0.01 mg/L Tungsten lamp with narrow band interference filter @ 575 nm Adaptation of the EPA method. The reaction between copper and the bicinchoninate reagent causes a purple tint in the sample.
• Remove the cuvet. • Add the content of one packet of HI 93702 reagent. Replace the cap and shake gently for about 15 seconds. • Reinsert the cuvet into the instrument. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 45 seconds and press READ DIRECT. In both cases “SIP” will blink during measurement. • The instrument directly displays concentration in mg/L of copper on the Liquid Crystal Display.
COPPER LOW RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0 to 1000 µg/L 1 µg/L ±10 µg/L ±5% of reading ±1 µg/L Tungsten lamp with narrow band interference filter @ 575 nm Adaptation of the EPA method. The reaction between copper and the bicinchoninate reagent causes a purple tint in the sample.
• Remove the cuvet. • Add the content of one packet of HI 93747 reagent. Replace the cap and shake gently for about 15 seconds. • Reinsert the cuvet into the instrument. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 45 seconds and press READ DIRECT. In both cases "SIP" will blink during measurement. • Multiply the reading on the Liquid Crystal Display by 10 to obtain the concetration in mg/L of oxygen demand.
CYANIDE SPECIFICATIONS Range Resolution Accuracy Typical EMC Dev. Light Source Method 0.000 to 0.200 mg/L 0.001 mg/L ±0.005 mg/L ±3% of reading ±0.001 mg/L Tungsten lamp with narrow band interference filter @ 610 nm Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Pyridine-Pyrazolone method. The reaction between cyanide and reagents causes a blue tint in the sample.
• Remove the cuvet and add 1 level spoon of HI 93714A Cyanide Reagent. Remember to close the reagent bottle immediately after use. Note: Pay attention to the way the spoon is filled: - do not press the powder; - do not overfill it. • Place the HDPE plastic stopper and cap immediately, to prevent the escape of chlorine gas which is developed during the reaction, and shake gently for 30 seconds.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 25 minutes and press READ DIRECT. In both cases "SIP" will blink during measurement. Note: Shake gently the cuvet 4 or 5 times during the first 20 minutes of the countdown prior to the measurement. Accuracy is not affected by undissolved reagent powder. • The instrument directly displays concentration in mg/L of cyanide on the Liquid Crystal Display.
CYANURIC ACID SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0 to 80 mg/L 1 mg/L ±1 mg/L ±15% of reading ±1 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the turbidimetric method. The reaction between cyanuric acid and the reagent causes a white suspension in the sample.
• Fill a graduated beaker up to the 25 mL mark with the sample, add the content of one packet of HI 93722 reagent and swirl gently to mix. • Fill a second cuvet with 10 mL of the reacted sample up to the mark. Replace the cap. 10 mL • Reinsert the cuvet into the instrument. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 45 seconds and press READ DIRECT. In both cases "SIP" will blink during measurement.
FLUORIDE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 2.00 mg/L 0.01 mg/L ±5% of reading ±0.01 mg/L Tungsten lamp with narrow band interference filter @ 575 nm Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, SPADNS method. The reaction between fluoride and the liquid reagent causes a red tint in the sample.
• Place the cuvet with the reacted distilled water (# 1) into the holder and ensure that the notch on the cap is positioned securely into the groove. #1 • Press TIMER and the display will show the countdown prior to zeroing the blank or, alternatively, wait for two minutes and press ZERO and "SIP" will blink on the display. • Wait for a few seconds and the display will show "-0.0-". Now the meter is zeroed and ready for measurement. • Remove the cuvet.
INTERFERENCES Negative interferences may be caused by: Alkalinity (as CaCO3) above 5000 mg/L Aluminum above 0.1 mg/L Iron, ferric above 10 mg/L Positive interferences may be caused by: Chloride above 700 mg/L Phosphate, ortho above 16 mg/L Sodium hexametaphosphate above 1.0 mg/L Sulfate above 200 mg/L Highly colored and turbid samples may require distillation Highly alkaline samples can be neutralized with nitric acid.
CALCIUM HARDNESS SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 2.70 mg/L 0.01 mg/L ±0.11 mg/L ±5% of reading ±0.01 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Calmagite method. The reaction between calcium and reagents causes a reddish-violet tint in the sample.
• Add 1 drop of HI 93720C EGTA solution to one cuvet (# 1), replace the cap and invert the cuvet several times to mix. This is the blank. #1 • Place the blank (# 1) into the holder and ensure that the notch on the cap is positioned securely into the groove. #1 • Press ZERO and "SIP" will blink on the display. • Wait for a few seconds and the display will show "-0.0-". Now the meter is zeroed and ready for measurement. • Remove the blank and insert the second cuvet (# 2) into the instrument.
Note: The test will detect any calcium contamination in the beaker, measuring syringes or sample cells. To test cleanliness, repeat the test multiple times until you obtain consistent results. Note: For better accuracy wash glassware with HCl 6N. SAMPLE DILUTION This meter is designed to determine low levels of hardness, typically found in water purification systems. When testing some other sources of water, it is not uncommon to come across levels of hardness that are greater than the range of this meter.
MAGNESIUM HARDNESS SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 2.00 mg/L 0.01 mg/L ±0.11 mg/L ±5% of reading ±0.02 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, EDTA colorimetric method. The reaction between magnesium and reagents causes a reddish-violet tint in the sample.
• Add 0.5 mL of HI 93719B Alkali solution and swirl to mix. Use this solution to rinse 2 cuvets. • Fill both cuvets up to the 10 mL mark. 10 mL 10 mL #1 #2 • Add 1 drop of HI 93719C EDTA solution to one cuvet (# 1), replace the cap and invert the cuvet several times to mix. This is the blank. #1 • Add 1 drop of HI 93719D EGTA solution to the second cuvet (# 2), replace the cap and invert the cuvet several times to mix. This is the sample.
• Remove the blank (# 1) and insert the sample (# 2) into the instrument, making sure that the notch on the cap is positioned securely into the groove. #2 • Press READ DIRECT. “SIP” will blink during measurement. • The instrument directly displays concentration in mg/L of magnesium hardness, as CaCO3, on the Liquid Crystal Display. • To convert the result to mg/L of magnesium (Mg), multiply by a factor of 0.243.
TOTAL HARDNESS HIGH RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 400 to 750 mg/L 5 mg/L ±10 mg/L ±2% of reading ±5 mg/L Tungsten lamp with narrow band interference filter @ 466 nm Adaptation of the EPA method 130.1. The reaction between calcium, magnesium and reagents causes a red-violet tint in the sample. REQUIRED REAGENTS Code HI 93735A-HR HI 93735B-0 HI 93735C-0 Description Quantity Hardness Indicator reagent HR 9.
• Add 2 drops of HI 93735B Buffer reagent. x2 • Replace the cap, shake gently to mix and wait for 15 seconds. • Place the cuvet into the holder and ensure that the notch on the cap is positioned securely into the groove. • Press ZERO and "SIP" will blink on the display. • Wait for a few seconds and the display will show "-0.0-". Now the meter is zeroed and ready for measurement. • Remove the cuvet and add the content of 1 packet of HI 93735C Fixing reagent.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 2 minutes and press READ DIRECT. In both cases "SIP" will blink during measurement. • The instrument directly displays the total hardness in mg/L of CaCO3, on the Liquid Crystal Display. • The conversion factors to convert readings in mg/L to French Degrees (FD), German Degrees (DD) and English Degrees (ED) of hardness are as follows: 1 mg/L= 0.1 FD = 0.0556 DD = 0.
TOTAL HARDNESS MEDIUM RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 200 to 500 mg/L 5 mg/L ±7 mg/L ±3% of reading ±5 mg/L Tungsten lamp with narrow band interference filter @ 466 nm Adaptation of the EPA method 130.1. The reaction between calcium, magnesium and reagents causes a red-violet tint in the sample. REQUIRED REAGENTS Code HI 93735A-MR HI 93735B-0 HI 93735C-0 Description Quantity Hardness Indicator reagent MR 9.
• Add 2 drops of HI 93735B Buffer reagent x2 • Replace the cap and shake gently to mix. Wait for 15 seconds. • Place the cuvet into the holder and ensure that the notch on the cap is positioned securely into the groove. • Press ZERO and "SIP" will blink on the display. • Wait for a few seconds and the display will show "-0.0-". Now the meter is zeroed and ready for measurement. • Remove the cuvet and add the content of 1 packet of HI 93735C Fixing reagent.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 2 minutes and press READ DIRECT. In both cases "SIP" will blink during measurement. • The instrument directly displays the hardness in mg/L of CaCO3, on the Liquid Crystal Display. • The conversion factors to convert readings in mg/L to French Degrees (FD), German Degrees (DD) and English Degrees (ED) of hardness are as follows: 1 mg/L= 0.1 FD = 0.0556 DD = 0.
TOTAL HARDNESS LOW RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0 to 250 mg/L 1 mg/L from 0 to 100 mg/L 5 mg/L from 100 to 250 mg/L ±5 mg/L ±4% of reading ±5 mg/L Tungsten lamp with narrow band interference filter @ 466 nm Adaptation of the EPA method 130.1. The reaction between calcium, magnesium and reagents causes a red-violet tint in the sample.
• Add 2 drops of HI 93735B-0 Buffer reagent. x2 • Replace the cap and shake gently to mix. Wait for 15 seconds. • Place the cuvet into the holder and ensure that the notch on the cap is positioned securely into the groove. • Press ZERO and "SIP" will blink on the display. • Wait for a few seconds and the display will show "-0.0-". Now the meter is zeroed and ready for measurement. • Remove the cuvet and add the content of 1 packet of HI 93735C Fixing reagent.
• Place the cuvet into the holder and ensure that the notch on the cap is positioned securely into the groove. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 2 minutes and press READ DIRECT. In both cases "SIP" will blink during measurement. • The instrument directly displays the hardness in mg/L of CaCO3, on the Liquid Crystal Display.
HYDRAZINE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0 to 400 µg/L 1 µg/L ±4% of full scale ±2 µg/L Tungsten lamp with narrow band interference filter @ 420 nm Adaptation of the ASTM Manual of Water and Environmental Technology, method D1385-88, p-Dimethylaminobenzaldehyde method. The reaction between hydrazine and the liquid reagent causes a yellow tint in the sample.
• Wait for a few seconds and the display will show a blinking "good". Now the meter is ready to take a zero measurement. • Fill a second cuvet with 10 mL of unreacted sample (up to the mark) and replace the cap. • Add 12 drops of the HI 93704 reagent to each cuvet. Replace the caps and shake gently to mix. x 12 #1 10 mL #2 x 12 #2 • Place the blank (#1) into the holder and ensure that the notch on the cap is positioned securely into the groove.
• Insert the cuvet with the reacted sample (# 2) into the instrument. Ensure that the notch on the cap is positioned securely into the groove. #2 • Press READ DIRECT and “SIP” will blink during measurement. • The instrument directly displays concentration in µg/L of hydrazine on the Liquid Crystal Display.
IODINE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.0 to 12.5 mg/L 0.1 mg/L ±0.1 mg/L ±5% of reading ±0.1 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, DPD method. The reaction between iodine and the reagent causes a pink tint in the sample.
• Wait for a few seconds and the display will show “-0.0-”. Now the meter is zeroed and ready for measurement. • Remove the cap and add the content of one packet of HI 93718 reagent. Replace the cap and shake gently for about 30 seconds to dissolve most of the reagent. • Reinsert the cuvet into the instrument. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 2 minutes and 30 seconds and press READ DIRECT.
IRON HIGH RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 5.00 mg/L 0.01 mg/L ±0.04 mg/L ±2% of reading ±0.01 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the EPA Phenantroline method 315B, for natural and treated waters. The reaction between iron and reagents causes an orange tint in the sample.
• Remove the cuvet and add the content of one packet of HI 93721 reagent. Replace the cap and shake until dissolution is complete. • Reinsert the cuvet into the instrument. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 3 minutes and press READ DIRECT. In both cases "SIP" will blink during measurement. • The instrument directly displays concentration in mg/L of iron on the Liquid Crystal Display.
IRON LOW RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0 to 400 µg/L 1 µg/L ±10 µg/L ±8% of reading ±1 µg/L Tungsten lamp with narrow band interference filter @ 575 nm Adaptation of the TPTZ Method. The reaction between iron and the reagent causes a violet tint in the sample.
• Place the cuvet into the holder and ensure that the notch on the cap is positioned securely into the groove. • Press ZERO and "SIP" will blink on the display. • Wait for a few seconds and the display will show "-0.0-". Now the meter is zeroed and ready for measurement. • Remove the cuvet. • Fill another graduated mixing cylinder up to the 25 mL mark with the sample. 25 ml 12 9 • Add the content of one packet of HI 93746 TPTZ reagent, close the cylinder and shake vigorously for 30 seconds.
• Insert the sample into the instrument. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 30 seconds and press READ DIRECT. In both cases "SIP" will blink during measurement. • The instrument directly displays concentration in µg/L of iron on the Liquid Crystal Display. INTERFERENCES Interference may be caused by: Cadmium above 4.0 mg/L Chromium3+ above 0.25 mg/L Chromium6+ above 1.2 mg/L Cobalt above 0.05 mg/L Copper above 0.
MANGANESE HIGH RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.0 to 20.0 mg/L 0.1 mg/L ±0.2 mg/L ±3% of reading ±0.1 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Periodate method. The reaction between manganese and reagents causes a pink tint in the sample.
• Wait for a few seconds and the display will show “-0.0-”. Now the meter is zeroed and ready for measurement. • Remove the cuvet. • Add the content of one packet of HI 93709A Citrate reagent. Replace the cap and shake gently until dissolution is complete. • Add the content of one packet of HI 93709B Sodium Periodate reagent. Replace the cap and shake gently until dissolution is complete. • Reinsert the cuvet into the instrument.
MANGANESE LOW RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0 to 300 µg/L 1 µg/L ±2 µg/L ±3% of reading ±1 µg/L Tungsten lamp with narrow band interference filter @ 575 nm Adaptation of the PAN Method. The reaction between manganese and the reagents causes an orange tint in the sample. REQUIRED REAGENT Code HI 93748A-0 HI 93748B-0 HI 93748C-0 HI 93703-51 Description Ascorbic acid Alkaline-cyanide sol. 0.1% PAN indicator Dispersing Agent Quantity 2 packets 0.
• Add 0.2 mL of the HI 93748B Alkaline-cyanide reagent solution to each cuvet, replace the caps and shake gently. • Add 1 mL of the HI 93748C 0.1% PAN indicator solution to each cuvet, replace the caps and shake gently. • Place the cuvet with the reacted deionized water (blank) into the holder and ensure that the notch on the cap is positioned securely into the groove. • Press TIMER and the display will show the countdown prior to zeroing the blank. Alternatively wait for 2 minutes and then press ZERO.
• Press READ DIRECT. "SIP" will blink during measurement. • The instrument directly displays concentration in µg/L of manganese on the display. Note: a temperature above 30°C may cause turbidity. In this case, before zeroing and taking readings, add 2-3 drops of Dispersing Agent (HI 93703-51) to each cuvet and swirl until complete dissolution of the turbidity.
MOLYBDENUM SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.0 to 40.0 mg/L 0.1 mg/L ±0.3 mg/L ±5% of reading ±0.1 mg/L Tungsten lamp with narrow band interference filter @ 420 nm Adaptation of the mercaptoacetic acid method. The reaction between molybdenum and the reagents causes a yellow tint in the sample.
• Wait for a few seconds and the display will show “-0.0-”. Now the meter is zeroed and ready for measurement. • Fill one graduated mixing cylinder up to the 25 mL mark with the sample. 25 mL • Add the content of one packet of HI 93730A molybdenum reagent, close the cylinder and invert it several times until dissolution is complete. • Add the content of one packet of HI 93730B molybdenum reagent to the cylinder, close and invert it several times until dissolution is complete.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for five minutes and press READ DIRECT. In both cases, "SIP" will blink during measurement. • The instrument directly displays concentration in mg/L of molybdenum on the Liquid Crystal Display.
NICKEL HIGH RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 7.00 g/L 0.01 g/L ±4% of reading ±0.02 g/L Tungsten lamp with narrow band interference filter @ 575 nm Adaptation of the photometric method. The reaction between nickel and the reagent causes a blue tint in the sample.
• Remove the cuvet and add the content of one packet of HI 93726 reagent. Replace the cap and shake gently until dissolution is complete. • Reinsert the cuvet into the instrument. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 1 minute and press READ DIRECT. In both cases "SIP" will blink during measurement. • The instrument directly displays concentration in g/L of nickel on the Liquid Crystal Display.
NICKEL LOW RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.000 to 1.000 mg/L 0.001 mg/L ±0.010 mg/L ±7% of reading ±0.001 mg/L Tungsten lamp with narrow band interference filter @ 575 nm Adaptation of the PAN method. The reaction between nickel and the reagents causes an orange tint in the sample. REQUIRED REAGENTS Code HI 93740A-0 HI 93740B-0 HI 93740C-0 HI 93703-51 Description Phthalate-phosphate 0.
• Add 1 mL of HI 93740B 0.3% PAN solution to each beaker, cap and swirl to mix. • Press TIMER and the display will show a countdown prior to adding reagent C or, alternatively, wait for 15 minutes. Then, in both cases, add one packet of HI 93740C EDTA reagent to each beaker, cap and swirl to mix until complete dissolution. • Fill one cuvet up to the mark with 10 mL of the blank. 10 mL • Place the cuvet into the holder and ensure that the notch on the cap is positioned securely into the groove.
• Press READ DIRECT and "SIP" will appear during measurement. • The instrument directly displays concentration in mg/L of nickel on the Liquid Crystal Display. Note: a temperature above 30°C may cause turbidity. In this case, before zeroing and taking readings, add 2-3 drops of Dispersing Agent (HI 93703-51) to each cuvet and swirl until complete dissolution of the turbidity.
NITRATE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.0 to 30.0 mg/L 0.1 mg/L ±0.5 mg/L ±10% of reading ±0.1 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the cadmium reduction method. The reaction between nitrate and the reagent causes an amber tint in the sample.
• Remove the cuvet and add the content of one packet of HI 93728 reagent. • Replace the cap and immediately shake vigorously for exactly 10 seconds by moving the cuvet up and down. Continue to mix by inverting the cuvet gently and slowly for 50 seconds, while taking care not to induce air bubbles. A deposit remains, but this does not affect the measurement. Time and way of shaking could sensitively affect the measurement. • Reinsert the cuvet into the instrument, taking care not to shake it.
NITRITE HIGH RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0 to 150 mg/L 1 mg/L ±4 mg/L ±4% of reading ±1 mg/L Tungsten lamp with narrow band interference filter @ 575 nm Adaptation of the Ferrous Sulfate method. The reaction between nitrite and the reagent causes a greenish-brown tint in the sample.
• Remove the cuvet. • Add the content of one packet of HI 93708 reagent. Replace the cap and shake gently until dissolution is complete. • Reinsert the cuvet into the instrument. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 10 minutes and press READ DIRECT. In both cases “SIP” will blink during measurement. • The instrument directly displays concentration in mg/L of nitrite on the Liquid Crystal Display.
NITRITE LOW RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 0.35 mg/L 0.01 mg/L ±0.02 mg/L ±4% of reading ±0.01 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the EPA Diazotization method 354.1. The reaction between nitrite and the reagent causes a pink tint in the sample.
• Remove the cuvet. • Add the content of one packet of HI 93707 reagent. Replace the cap and shake gently for about 15 seconds. • Reinsert the cuvet into the instrument. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 6 minutes and press READ DIRECT. In both cases “SIP” will blink during measurement. • The instrument directly displays concentration in mg/L of nitrite nitrogen.
CHEMICAL OXYGEN DEMAND HIGH RANGE SPECIFICATIONS Range Resolution Precision Typical EMC Deviation Light Source Method 0 to 15000 mg/L 10 mg/L (1 mg/L x 10) Standard deviation ±220 @ 10000 mg/L ±1 mg/L Tungsten lamp with narrow band interference filter @ 610 nm Adaptation of the EPA method 410.4. Oxidizable organic compounds reduce the dichromate ion (orange) to the Chromium (III) ion (green). The amount of Chromium (III) formed is determined.
• For sample digestion use a block heater reactor with holes to accommodate digestion vials. Use of the optional HI 740217 safety shield is strongly recommended. Preheat the Hanna Reactor C 9800 to 150 °C (302°F). For correct use of the reactor follow Reactor Instruction Manual. Do not use an oven or microwave because leaking samples can generate a corrosive and possibly explosive atmosphere. • Remove the cap from a reagent vial for the COD-HR.
• Leave the vials in the tube rack to cool to room temperature. Do not shake or invert them anymore otherwise the samples may become turbid. • Select the program number corresponding to Oxygen Demand, Chemical (COD) - HR on the secondary LCD by pressing PROGRAM W and V. • Place carefully the COD-vial adapter in the instrument and check that it fits properly. • Place the blank into the holder. • Press ZERO and “SIP” will blink on the display. • Wait for a few seconds and the display will show “-0.0-”.
CHEMICAL OXYGEN DEMAND MEDIUM RANGE SPECIFICATIONS Range Resolution Precision Typical EMC Deviation Light Source Method 0 to 1500 mg/L 1 mg/L Standard deviation ±22 @ 1000 mg/L ±1 mg/L Tungsten lamp with narrow band interference filter @ 610 nm Adaptation of the EPA method 410.4. Oxidizable organic compounds reduce the dichromate ion (orange) to the Chromium (III) ion (green). The amount of Chromium (III) formed is determined.
• For sample digestion use a block heater reactor with holes to accommodate digestion vials. Use of the optional HI 740217 safety shield is strongly recommended. Preheat the Hanna Reactor C 9800 to 150 °C (302°F). For correct use of the reactor follow Reactor Instruction Manual. Do not use an oven or microwave because leaking samples can generate a corrosive and possibly explosive atmosphere. • Remove the cap from a reagent vial for the COD-MR.
• Leave the vials in the tube rack to cool to room temperature. Do not shake or invert them anymore otherwise the samples may become turbid. • Select the program number corresponding to Oxygen Demand, Chemical (COD) - MR on the secondary LCD by pressing PROGRAM W and V. • Place carefully the COD-vial adaptor in the instrument and check that it fits properly. • Place the blank into the holder. • Press ZERO and “SIP” will blink on the display. • Wait for a few seconds and the display will show “-0.0-”.
CHEMICAL OXYGEN DEMAND LOW RANGE SPECIFICATIONS Range Resolution Precision Typical EMC Deviation Light Source Method 0 to 150 mg/L 1 mg/L Standard deviation ±4 @ 150 mg/L ±1 mg/L Tungsten lamp with narrow band interference filter @ 420 nm Adaptation of the EPA method 410.4. Oxidizable organic compounds reduce the dichromate ion (orange) to the Chromium (III) ion (green). The amount of remaining dichromate is determined.
• For sample digestion use a block heater reactor with holes to accommodate digestion vials. Use of the optional HI 740217 safety shield is strongly recommended. Preheat the Hanna Reactor C 9800 to 150 °C (302°F). For correct use of the reactor follow Reactor Instruction Manual. Do not use an oven or microwave because leaking samples can generate a corrosive and possibly explosive atmosphere. • Remove the cap from a reagent vial for the COD-LR.
• Leave the vials in the tube rack to cool to room temperature. Do not shake or invert them anymore otherwise the samples may become turbid. • Select the program number corresponding to Oxygen Demand, Chemical (COD) - LR on the secondary LCD by pressing PROGRAM W and V. • Place carefully the COD-vial adapter in the instrument and check that it fits properly. • Place the blank into the holder. • Press ZERO and “SIP” will blink on the display. • Wait for a few seconds and the display will show “-0.0-”.
DISSOLVED OXYGEN SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.0 to 10.0 mg/L 0.1 mg/L ±0.4 mg/L ±3% of reading ± 0.1 mg/L Tungsten lamp with narrow band interference filter @ 420 nm Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Azide modified Winkler method. The reaction between dissolved oxygen and the reagents causes a yellow tint in the sample.
• Add some more sample, to fill the bottle completely. Replace the cap again and ensure that a part of the sample spills over. This is to make sure that no air bubbles have been trapped inside, which would corrupt the reading. • Invert several times the bottle. The sample becomes orange-yellow and a flocculant agent will appear. • Let the sample stand and the flocculant agent will start to settle. • After approximately 2 minutes, when the upper half of the bottle becomes limpid, add 10 drops of HI 93732C.
• Wait for a few seconds and the display will show “-0.0-”. Now the meter is zeroed and ready for measurement. • Remove the cuvet. • Fill another cuvet up to the mark with 10 mL of the reacted sample and replace the cap. • Reinsert the cuvet into the instrument. • Press READ DIRECT and "SIP" will blink during measurement. • The instrument will then directly display the concentration of dissolved oxygen in mg/L on the Liquid Crystal Display.
pH SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 6.5 to 8.5 pH 0.1 pH ±0.1 pH ±0.1 pH Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the Phenol Red method. The reaction with the reagent causes a yellow to red tint in the sample. REQUIRED REAGENTS Code HI 93710-0 Description Phenol Red Indicator Quantity 5 drops REAGENT SETS HI 93710-01 Reagents for 100 pH tests HI 93710-03 Reagents for 300 pH tests For other accessories see page 141.
• Remove the cuvet and add 5 drops of HI 93710 Phenol Red Indicator. Replace the cap and mix the solution. x5 • Reinsert the cuvet into the instrument. • Press the READ DIRECT key and “SIP” will blink on the display during measurement. • The instrument directly displays the pH measured value on the Liquid Crystal Display.
PHOSPHATE HIGH RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Dev. Light Source Method 0.0 to 30.0 mg/L 0.1 mg/L ±1 mg/L ±4% of reading ±0.1 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Amino Acid method. The reaction between phosphate and reagents causes a blue tint in the sample.
• Press ZERO and “SIP” will blink on the display. • Wait for a few seconds and the display will show "-0.0-". Now the meter is zeroed and ready for measurement. • Remove the cuvet. • Add 10 drops of HI 93717A x 10 Molybdate reagent. • Add to the cuvet 0.5 mL of the solution prepared in the HI 93717B bottle. Replace the cap and mix the solution. • Reinsert the cuvet into the instrument.
PHOSPHATE LOW RANGE SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 2.50 mg/L 0.01 mg/L ±0.04 mg/L ±4% of reading ±0.01 mg/L Tungsten lamp with narrow band interference filter @ 610 nm Adaptation of the Ascorbic Acid method. The reaction between phosphate and the reagent causes a blue tint in the sample.
• Wait for a few seconds and the display will show “-0.0-”. Now the meter is zeroed and ready for measurement. • Remove the cuvet and add the content of one packet of HI 93713 reagent. Replace the cap and shake gently (for about 2 minutes) until the powder is completely dissolved. • Reinsert the cuvet into the instrument. • Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 3 minutes and press READ DIRECT.
PHOSPHORUS SPECIFICATIONS Range Resolution Accuracy Typical EMC dev. Light Source Method 0.0 to 15.0 mg/L 0.1 mg/L ±0.3 mg/L ±4% of reading ±0.2 mg/L Tungsten lamp with narrow band interference filter @ 525 nm Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Amino Acid method. The reaction between phosphorus and the reagents causes a blue tint in the sample.
• Wait for a few seconds and the display will show "-0.0-". Now the meter is zeroed and ready for measurement. • Remove the cuvet. • Add 10 drops of HI 93706A Molybdate reagent. x 10 • Add the content of one packet of HI 93706B-0 Phosphorus Reagent B (Amino Acid) to the cuvet. Replace the cap and shake gently until dissolution is complete. • Reinsert the cuvet into the instrument.
SILICA SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 2.00 mg/L 0.01 mg/L ±0.03 mg/L ±3% of reading ±0.01 mg/L Tungsten lamp with narrow band interference filter @ 610 nm Adaptation of the ASTM Manual of Water and Environmental Technology, D859, Heteropoly Blue method. The reaction between silica and reagents causes a blue tint in the sample.
• Wait for 1 minute. This is the blank. 1’ 12 9 3 6 • Place the cuvet into the holder and ensure that the notch on the cap is positioned securely into the groove. • Press ZERO and “SIP” will blink on the display. • Wait for a few seconds and the display will show “-0.0-”. Now the meter is zeroed and ready for measurement. • Remove the cuvet and add the content of one packet of HI 93705C Amino acid reagent and shake until it is completely dissolved. • Reinsert the cuvet into the instrument.
INTERFERENCES Interference may be caused by: Phosphate above 60 mg/L (causes a 2% reduction in reading) Phosphate above 75 mg/L (causes an 11% reduction in reading) Sulfide and high concentration of iron Eliminate color and turbidity interferences by zeroing the meter with the original water sample.
SILVER SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.000 to 1.000 mg/L 0.001 mg/L ±0.005 mg/L ±10% of reading ± 0.001 mg/L Tungsten lamp with narrow band interference filter @ 575 nm. Adaptation of the PAN method. The reaction between silver and reagents causes an orange tint in the sample.
• Add exactly 1.0 mL of HI 93737B reagent to the other beaker (the sample) and swirl gently to mix. #2 • Press TIMER and the display will show the countdown prior to adding reagent C or, alternatively, wait for 2 minutes. Then, in both cases, add exactly 1.0 mL of HI 93737C reagent to each beaker and swirl. #2 #1 • Press TIMER or, alternatively, wait for 2 minutes. Then, in both cases, add 1.0 mL of HI 93737D reagent to each beaker and swirl.
• Press ZERO and "SIP" will blink on the display. • Wait for a few seconds and the display will show "-0.0.-". Now the meter is zeroed and ready for measurement. 10 mL • Fill a second cuvet up to the mark with 10 mL of the reacted sample. #2 #2 • Insert the second cuvet into the instrument. • Press READ DIRECT and the display will show "SIP" during measurement. • The instrument directly displays concentration in mg/L of silver on the Liquid Crystal Display.
ZINC SPECIFICATIONS Range Resolution Accuracy Typical EMC Deviation Light Source Method 0.00 to 3.00 mg/L 0.01 mg/L ±0.03 mg/L ±3% of reading ±0.01 mg/L Tungsten lamp with narrow band interference filter @ 575 nm Adaptation of the Standard Methods for the Examination of Water and Wastewater, 18th edition, Zincon method. The reaction between zinc and the reagents causes an orange to a dark violet tint in the sample.
• Fill one cuvet with 10 mL of the reacted sample up to the mark. 10 mL • Place the cap and insert the cuvet into the cell and ensure that the notch on the cap is positioned securely into the groove. • Press ZERO and “SIP” will blink on the display. • Wait for a few seconds and the display will show “-0.0-”. Now the meter is zeroed and ready for measurement. • Remove the cuvet and add 0.5 mL of HI 93731B cyclohexanone to the cuvet.
• Press TIMER and the display will show the countdown prior to the measurement or, alternatively, wait for 3 minutes and 30 seconds and press READ DIRECT. In both cases "SIP" will blink during measurement. • The instrument directly displays concentration in mg/L of zinc on the Liquid Crystal Display. INTERFERENCES Interference may be caused by: Aluminum above 6 mg/L Cadmium above 0.
INTERFACE WITH PC To connect your meter to the PC use the optional HI 920010 (available from your Hanna Dealer). Make sure that your meter is switched off and plug the connectors, one into the meter RS 232C socket, the other into the serial port of your PC. Note: Cables other than HI 920010 may use a different configuration, in which case, communication between the meter and the PC may not be possible.
/PUP - Program Up /PDN - Program Down /PTM - Turn Test Mode On /Brx - Set the baud rate 1 - 300 2 - 600 3 - 1200 4 - 2400 /KBL - Lock Keyboard /KBU - Unlock Keyboard ?PR# - Send Current Program Number ?BRQ - Send current baud rate 1 - 150 2 - 300 3 - 600 4 - 1200 5 - 2400 ?CNQ - Send Concentration (three bytes) conc decimal point unit m - ppm b - ppb t - ppt u - pcu h - pH ?ERR - send error / status information 0 - No error 1 - CAP 2 - HI 3 - ZERO 4 - LO 5 - IDLE 6 - ZERO DONE 7 - TIMED READ DROPS/ m L CRO
STANDARD METHODS Description Range Method Aluminum Ammonia, HR Ammonia, MR Ammonia, LR Bromine Chlorine, Free Chlorine, Free HR Chlorine, Total Chlorine, Total HR Chlorine Dioxide Chromium VI, HR Chromium VI, LR COD HR COD MR COD LR Color of Water Copper, HR Copper, LR Cyanide Cyanuric Acid Fluoride Hardness, Ca Hardness, Mg Hardness, Total HR Hardness, Total MR Hardness, Total LR Hydrazine Iodine Iron, HR Iron, LR Manganese, HR Manganese, LR Molybdenum Nickel, HR Nickel, LR Nitrate Nitrite, HR Nitrite,
BATTERY REPLACEMENT Besides to the 12 VDC supply, these meters can also be fed by two 9V batteries. To prolong battery life, switch your meter off after use. However, the meter has an auto-shut off feature that will turn itself off after 10 minutes of non-use. A blinking "LOBAT" on the display, while a measurement is in progress, indicates low voltage and the batteries should be replaced.
ACCESSORIES REAGENT SETS HI 93700-01 100 ammonia LR tests HI 93700-03 300 ammonia LR tests HI 93701-01 100 free chlorine tests (powder) HI 93701-03 300 free chlorine tests (powder) HI 93701-F 300 free chlorine tests (liquid) HI 93701-T 300 total chlorine tests (liquid) HI 93702-01 100 copper HR tests HI 93702-03 300 copper HR tests HI 93704-01 100 hydrazine tests HI 93704-03 300 hydrazine tests HI 93705-01 100 silica tests HI 93705-03 300 silica tests HI 93706-01 100 phosphorus tests HI 93706-03 300 phospho
HI 93723-01 100 chromium VI HR tests HI 93723-03 300 chromium VI HR tests HI 93726-01 100 nickel HR tests HI 93726-03 300 nickel HR tests HI 93728-01 100 nitrate tests HI 93728-03 300 nitrate tests HI 93729-01 100 fluoride tests HI 93729-03 300 fluoride tests HI 93730-01 100 molybdenum tests HI 93730-03 300 molybdenum tests HI 93731-01 100 zinc tests HI 93731-03 300 zinc tests HI 93732-01 100 dissolved oxygen tests HI 93732-03 300 dissolved oxygen tests HI 93733-01 100 ammonia HR tests HI 93733-03 300 ammon
HI 740142 HI 740143 HI 740144 HI 740157 HI 740219 HI 740220 HI 92000 HI 920010 HI 93703-50 1 mL graduated syringe 1 mL graduated syringe (6 pcs) Pipette tip (6 pcs) Plastic refilling pipette (20 pcs) COD adapter 25 mL glass cylinders with caps (2 pcs) Windows® Compatible Software PC connection cable Cuvets cleaning solution (230 mL) WARRANTY All Hanna Instruments meters are warranted for two years against defects in workmanship and materials when used for their intended purpose and maintained according to
CE DECLARATION OF CONFORMITY Recommendations for Users Before using these products, make sure that they are entirely suitable for the environment in which they are used. Operation of these instruments in residential area could cause unacceptable interferences to radio and TV equipments, requiring the operator to take all necessary steps to correct interferences. Any variation introduced by the user to the supplied equipment may degrade the instruments' EMC performance.
HANNA LITERATURE Hanna publishes a wide range of catalogs and handbooks for an equally wide range of applications. The reference literature currently covers areas such as: • • • • • • • Water Treatment Process Swimming Pools Agriculture Food Laboratory Thermometry and many others. New reference material is constantly being added to the library. For these and other catalogs, handbooks and leaflets contact your dealer or the Hanna Customer Service Center nearest to you.
USER NOTES 146
MANC200R4 08/03 W W W . H A N N A 148 I N S T .