User Manual

24
At the end of the measurement, the third calibration
point (100 NTU) is displayed on the primary LCD,
“CAL P.3” on the secondary LCD, and “READ” tag
will blink.
At this moment it is possible to exit calibration
by pressing CAL. The instrument will memorize
the two-point calibration data and will return
to measurement mode.
THREE-POINT CALIBRATION
To perform a three-point calibration, continue the
procedure with the following steps:
Remove the second standard cuvet.
Place the 100 NTU standard cuvet (or the
prepared formazin standard) into the holder,
with the cuvet mark aligned with the sign on
the instrument.
Close the lid and press READ
. The display
will show blinking dashes and the icons for
cuvet, detectors and lamp will appear during
measurement.
At the end of the measurement, the fourth
calibration point (750 NTU) is displayed on the
primary LCD and “CAL P.4” on the secondary LCD.
At this moment it is possible to exit calibration
by pressing CAL. The instrument will memorize
the three-point calibration data and will return
to measurement mode.
FOUR-POINT CALIBRATION
To perform a four-point calibration, continue the
procedure with the following steps:
Remove the third standard cuvet.
Place the 750 NTU standard cuvet (or the prepared
formazin standard) into the holder, with the cuvet
mark aligned with the sign on the instrument.