Safety Information

Control solution testing
Understanding out-of-range control test results
Compare the result displayed on the meter
to the control solution range printed on the
test strip vial. Each control solution have a
different control solution range.
Out-of-range results may be due to:
not following the instructions detailed in steps 1–6
expired or contaminated control solution
expired or damaged test strip
a problem with the meter.
CAUTION: The control solution range printed on the test strip
vial is for Sejoy Control Solution only. It is not a recommended
range for your blood glucose level.
CAUTION: If you continue to get control solution test results
that fall outside the range printed on the test strip vial, Do Not use
the meter, the test strips, or the control solution. Contact Customer
33
use of a test strip or control solution that passed its discard date
CTRL
mmol/L
mg/dl
CTRL
mmol/L
mg/dl
CTRL
mmol/L
mg/dl
3-5
54- 90
7.5- 12.5
135- 225
12- 20
216- 360
Service. Contact us at (+86)571-81957767.