Safety Information

Introduction
Thank you for choosing Sejoy BG-210b blood glucose monitoring
system. Before you start testing, carefully read this .Owner’s Booklet
Intended Use
Test Principle
1
Peel off the insulation film from battery compartment before first use.
*
The strip measures glucose by using amperometric technology
employing a glucose oxidase reaction.When whole blood or con-trol
solution is drawn into the tip of a test strip,glucose in the sample reacts
with chemicals and produces an electrical current. The meter
measures electrical current and calculates amount of glucose. The
glucose result is displayed as a calculated plasma value.
Venous whole samples collection and preparation should be obtained
by healthcare professionals.
The Sejoy BS-302 Test Strips are used with the Sejoy BG-210b
Glucose Meters to quantitative measuring blood glucose with fresh
capillary whole blood from the finger or venous blood.The system is
intended for use outside the body (in vitro diagnostic use) by people
with diabetes at home and by health care professional in a clinical
setting as an aid to monitor theeffectiveness of diabetes control.
It should not be used for the diagnosis.
Blood