Manual Zelos Art. No. 0974 FB 04-247 Rev. 3 Revision: 16.11.
Directory Foreword 3 Safety instructions Cleaning & Disinfection 4 6 Getting to know the unit 7 Before the use Range of application Installation- site 8 8 9 Application Operation Working positioning Suction Changing filter bag Changing control unit filter Clamping the bur/ tool Setting up the speed Memory keys M1-M3 Button on the handpiece Foot switch LCD Display Messages 10 10 11 13 13 13 14 14 12 14 15 15 16 Maintenance & Care Cleaning the chuck Handling instructions 17 17 18 Troubleshooting
Foreword The Zelos foot care unit was produced according to strict quality criteria and uses the latest technology in its design. The images in this manual may vary slightly from the product you have bought as we are constantly developing our products. This user manual includes a detailed description and explanation of how to use the Zelos unit. If you have any questions or suggestions you can contact us by telephone or email. This user manual belongs with the Zelos unit. Keep it somewhere safe.
Safety Instructions The equipment should only be used by trained operators. Prevent your hair from wrapping itself around moving parts. If applicable, wear a hair net. The installation of the working area has to correspond to the relevant regulations. Set up the unit in such a way, that the air slits are not blocked and the extracted air can escape well. When working with materials, which might create dust or moisture, use a drilling unit with suction or spray tech-nique.
Safety Instructions liquids). the inserted tool sits properly in the chuck and will not cant/jam while clamping it the tool is approved for using it at the max speed (see manufacturer´s instructions), or only use the speed required by the instrument the tool is absolutely dry. Always follow the operating instructions. Do not carry out any repairs, modifications or maintenance work yourself. This is only to be accomplished by an authorized specialist.
Safety Instructions Cleaning and disinfection The device is designed to be compatible with rotating instruments which do not penetrate the skin or mucosal or inadvertently may still cause injuries. For this application please adhere to the cleaning and disinfection instructions provided below. After every treatment please carry out the following steps: Remove the rotating tool (e.g. bur, capper, etc.) first. Then handle the tool according to the instructions of it`s manufacturer.
Getting to know the unit 1 2 3lernen 4 5 6 Gerät kennen 11 12 13 7 8 9 14 1 Power switch (at the side of the unit) 2 Reduce speed 3 Increase speed 4 Change direction of speed 5 Display for speed [rpm] 6 Display for suction level, time, change of the filter bag, etc.
Before the use Please read through the instructions carefully before using the unit and carefully note the safety advice. Always keep this manual accessible to any users of the drill. Operation The working area must conform to the conditions described. Set up the unit in such a way, that the air slits are not blocked and the extracted air can escape well. To guarantee unhindered air flow there must a gap of at least 12cm to the right of the device.
Before the use Should the handpiece be twisted by the hose you can adjust that at the control unit. . Check that the filter bag and dust filter are correctly fitted. (See page 13) Ensure that the tool is fixed (See page 14). On the back of the device you will find the main power switch. Press this switch to turn the unit on. To use the handpiece and suction function press the button (14) on the front of the unit. The word "Language?" will appear on the display.
Use custom settings simply press M1. Other values can be saved using buttons M2 and M3. Operation In the following section the order in which you should use the device is described. Additionally the individual steps will be explained in detail. Read the complete manual before use. After choosing a rotation speed and suction power you can start working. If you want to take a break or finish the job press the standby button (14). The handpiece motor and suction will turn off.
Use Work positioning The following images are examples of how to work on various parts of the foot. Always make sure that the foot is firmly gripped and that you can work freely and without strain.
Use 12
Use Suction Switch on the suction with button (8). The suction output can be regulated in 5 levels. You will see the selected level in the display. Level 0 would mean that the suction is turned off. Pull out the lid of the dust chamber and remove the old filter. In case the suction motor overheats, the unit will switch off automatically. Leave the unit to cool off for a while before starting it again. Should the same failure appear again, please return the unit to your supplier for inspection.
Use Clamping the tool/ bur Memory button M1-M3 The handpiece is equipped with a chuck, which can hold all burs with a diameter of 2.35 mm (this is the diameter usually used for tools in podiatry). To open the chuck, push the knob forward with your thumb and leave it there. Now insert the tool as far as possible. Pull back the knob and the tool will be firmly clamped Frequently used speed or suction levels can be saved. Therefore three memory buttons are available.
Use By pressing the button, you will access the selective list for settings of the device. With the buttons <> and <> you can go through the list and select a setting by pressing . With you can stop a function to return to the main menu. Footswitch (0970) Operation of the footswitch: Only use the original hadewe foot switch, to avoid any damages. Insert the plug of the foot switch into the socket with the symbol .
Use on> or >light off< Press . Messages Check the filter bag see page 8 Display date of Maintenance Select from the list Approve with the button . The date of the last service will appear. If you send in the device for maintenance, hadewe or an approved service agent will update the date of service for you. Handpiece check No handpiece is connected or the handpiece is faulty. Maintenance recommended - After longer operation of the unit (approx.
Maintenance & Care Set the screw spanner (wrench) on the axle and hold it tight. Now use the front of the second spanner to unscrew the chuck. Cleaning the chuck Because of the additional protection washer (see right), not much dirt is usually entering the handpiece. In the process of time, some dirt might collect itself within the handpiece, and might affect its functions. Therefore the chuck has to be cleaned monthly.
Maintenance & Care handpiece, thorough inspections and examinations always have proved that the rust had a foreign origin (which was usually caused by the tools used in these circumstances).
Maintenance & Care Working with high pressure will not improve the outcome of your work, since the motor will even be slowed down. A second disadvantage is the stronger impact on the ball bearings and the motor, which will shorten their life span in the long run. In case of sending the device by mail remove the dust filter before. Only suck up dust or milling waste, never suck up larger particles such as small bits of cotton wool as this may be get caught in the handpiece and lead to blockages.
Troubleshooting Handpiece vibrates, is very noisy and/ or becomes warm in the front part Use of bent burs/ tools Use a different bur/ tool to test this. The bur/tool itself can be tested with the hadewe Bur Shaft Tester (Art. No. (4990) to see if it is bent). Unit does not work at all. Check, if the main switch of the unit is turned on at the side of the unit.
Technical Details Operating time 0974 Zelos Total weight: 4,1 kg Dimension: W273xH142xD186mm Input:230V, AC voltage 50Hz, 1A vacuum: 2,9 – 5,8 kPa Fuse primary: F1+F3: 3,15A time delayed safety fuse F3: 400 mA time delayed safety fuse Fuse Sec.: Thermal fuse Protection class II Accuracy of speed display: ±10% Handpiece: Type B applied part Diameter: 17-22 mm Length: 163mm Speed: 6.000 - 22.000 rpm The device is designed for extended usage.
Technical Details mask for your face according EN 14683 version IIR or EN 149 FFP2. Accessories Art. No. 5115 5179 3752 0970 Disposal of dust filter and coarse dust filter The filter bag and the coarse dust filter are considered low contaminated waste. Dispose those in intransparent, liquid- tight, resistant plasticks sacks and to not dispose them compressed in the regular houshold garbage.
Technical Details Symbol Safety Check Caution! A safety check must be done in compliance with the regulations of the country where the device is used. The inspection interval is once a year.
Electromagnetic Compatibility Guidance and Manufacturer's Declaration – Electromagnetic Emissions The device is suitable for use in the specified electromagnetic environment. The customer and/or the user of the pa-on should assure that it is used in an electromagnetic environment as described below: Emissions Test Compliance Electromagnetic Environment – Guidance RF emissions according to CISPR Group 1 The device uses RF energy 11 only for its internal function.
Electromagnetic Compatibility Guidance and Manufacturer's Declaration – Electromagnetic Immunity IEC 60601-1-2 The device is suitable for use in the specified electromagnetic environment.
Electromagnetic Compatibility Guidance and Manufacturer's Declaration – Electromagnetic Immunity IEC 60601-1-2 The device is suitable for use in the specified electromagnetic environment.
Electromagnetic Compatibility broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered. If the measured field strength in the location in which the device is used exceeds the applicable RF compliance level above, the device should be observed to verify normal operation.
hadewe GmbH Grambartstraße 10 30165 Hannover Germany info@hadewe.de www.hadewe.