Index of contents Prologue 3 Safety instructions Cleaning & Disinfection 4 6 Getting to know the unit 7 Before the use Range of application Installation- site 8 8 8 Application Operation Working positioning Suction Changing filter bag Changing control unit filter Setting up the speed Foot switch 10 10 11 13 13 13 14 14 Handling Details 15 Troubleshooting 17 Technical data 18 2
Prologue The Helius foot care unit was produced according to strict quality criteria and uses the latest technology in its design. The images in this manual may vary slightly from the product you have bought as we are constantly developing our products. This user manual includes a detailed description and explanation of how to use the Helius unit. If you have any questions or suggestions you can contact us by telephone or email. This user manual belongs with the Helius unit. Keep it somewhere safe.
Safety instructions The equipment should only be used by trained operators. Prevent your hair from wrapping itself around moving parts. If applicable, wear a hair net. The installation of the working area has to correspond to the relevant regulations. Set up the unit in such a way, that the air slits are not blocked and the extracted air can escape well. When working with materials, which might create dust or moisture, use a drilling unit with suction or spray tech-nique.
Safety instructions liquids). the inserted tool sits properly in the chuck and will not cant/jam while clamping it the tool is approved for using it at the max speed (see manufacturer´s instructions), or only use the speed required by the instrument the tool is absolutely dry. Always follow the operating instructions. Do not carry out any repairs, modifications or maintenance work yourself. This is only to be accomplished by an authorized specialist.
Safety instructions Cleaning and disinfection The device is designed to be compatible with rotating instruments which do not penetrate the skin or mucosal or inadvertently may still cause injuries. For this application please adhere to the cleaning and disinfection instructions provided below. After every treatment please carry out the following steps: Remove the rotating tool (e.g. bur, capper, etc.) first. Then handle the tool according to the instructions of it`s manufacturer.
Getting to know the unit 1 the unit 2 To get to know 3 4 8 1 Decrease the suction output 2 Increase the suction output 3 Reduce the speed 4 Increase the speed 5 Display of speed/ suction level [rpm] 6 Change direction of rotation 7 LED: ON during reverse rotation 8 Switching ON/OFF the unit (Standby) 9 Connecting socket for foot switch 7 5 6 9 7
Before the use Please read through the instructions carefully before using the unit and carefully note the safety advice. Always keep this manual accessible to any users of the drill. Installation site The working area must conform to the conditions described. Set up the unit in such a way, that the air slits are not blocked and the extracted air can escape well. To guarantee unhindered air flow there must a gap of at least 12cm to the right of the device.
Before the use . Check that the filter bag and dust filter are correctly fitted. Ensure that the tool is fixed. On the left side in the back of the device you will find the main power switch. Press this switch to turn the unit on. The device is now on standby Pick up the handpiece. To use the handpiece and suction function press the button (8) on the front of the unit.
Application If you want to take a break or finish the job press the standby button (9). The handpiece motor and suction will turn off. The handpiece can now be placed in its holder on the right of the unit. Alternatively you can switch the device to standby by pressing the button on the foot switch. Operation In the following section the order in which you should use the device is described. Additionally the individual steps will be explained in detail. Read the complete manual before use.
Application Work positioning The following images are examples of how to work on various parts of the foot. Always make sure that the foot is firmly gripped and that you can work freely and without strain.
Application 12
Application Suction Switch on the suction with button (2). The suction output can be regulated in 5 levels. You will see the selected level in the display. Level 0 would mean that the suction is turned off. In case the suction motor overheats, it will switch off automatically. Leave the unit to cool off for a while before starting it again. Should the same failure appear again, please return the unit to your supplier for inspection. Pull out the lid of the dust chamber and remove the old filter.
Application Exchange control unit filter When the device is brought out of standby mode it will run with the rotation speed and suction power used last. Take care to not start the mill at an overly high rotation speed. Select the required rotation speed before changing the milling tool. This filter (Art. No. 3752) has to be exchanged after every 200 operating hours. Remove the old filter and place the new one in its holder (see picture). Foot switch (Art.
Handling details Never put the handpiece into any kind of liquid, and never oil or grease the handpiece. Clean its surface with a damped cloth. Pay attention, that no moisture or liquid enters the handpiece. Never use spray dis-infection. remove the bur/ tool from it. Always make sure that you do not let the handpiece fall down or exposes it to any shocks. The ballbearings and the motor could be damaged by this. Never put any wet burs or instruments into the handpiece.
Handling details the dust filter before. Only suck up dust or milling waste, never suck up larger particles such as small bits of cotton wool as this may be get caught in the handpiece and lead to blockages. Always working with maximum speed might have a similar effect. Please make sure that the speed setting you use is according to the work you are doing and the instruments you are working with.
Troubleshooting Handpiece vibrates, is very noisy and/ or becomes warm in the front part Use of bent burs/ tools Use a different bur/ tool to test this. The bur/tool itself can be tested with the hadewe Bur Shaft Tester (Item. No. 4990) to see, if it is bent. Maximum allowed speed of the tool was exceeded, which caused a stronger vibration pay attention to the tool (i.g.
Technical Details 0965 Helius R Weight: 4,4kg Dimensions:W 257x H 140x D 257mm Input: 230/240 AC Voltage.,50Hz Output: 550W vacuum: 2,9 – 5,8 kPa Fuse primary: F1: 3,15 A time delayed, safety fuse F3: 315 mA time delayed safety fuse Fuse Sec.: Thermal fuse Protection class I Accuracy of speed display: ±10% Handpiece: Type B applied part Diameter: 18-25mm Length: 130mm Speed: 6.000-30.
Technical Details Compulsory registration Operators, distributors and retailers who were informed about the occurrences regarding §29 of the Medical Devices law must report this. The method of reporting is detailed in the Medical Devices Safety Plan Ordinance. Outside of Germany the corresponding local conditions apply Disposal The device and filter may contain infectious material such as residual sanding dust or skin particles. Therefore please bear in mind the following disposal instructions.
Technical Details Optical control, unit discharge current <500µA, discharge current at applied part <500µA. Functional inspection of display, on/off-button, suction/spray, speed and clamping system. Safety and function deficiencies have not been found. By providing this information an inspection before use can be renounced as per DIN EN 62535, which otherwise would have to be ordered by the user. Accessories Art. Nr.
Electromagnetic Compatibility Guidance and Manufacturer's Declaration – Electromagnetic Emissions The device is suitable for use in the specified electromagnetic environment. The customer and/or the user of the pa-on should assure that it is used in an electromagnetic environment as described below: Emissions Test Compliance Electromagnetic Environment – Guidance RF emissions according to CISPR Group 1 The device uses RF energy 11 only for its internal function.
Electromagnetic Compatibility Guidance and Manufacturer's Declaration – Electromagnetic Immunity IEC 60601-1-2 The device is suitable for use in the specified electromagnetic environment.
Electromagnetic Compatibility Guidance and Manufacturer's Declaration – Electromagnetic Immunity IEC 60601-1-2 The device is suitable for use in the specified electromagnetic environment.
Electromagnetic Compatibility a) Field strengths from fixed transmitters, such as base stations for radio (cellular/cordless) telephones and land mobile radios, amateur radio, AM and FM radio broadcast, and TV broadcast cannot be predicted theoretically with accuracy. To assess the electromagnetic environment due to fixed RF transmitters, an electromagnetic site survey should be considered.