LTE tele-health modem Instruction (H3G-900) H3 System Co.
< Revision Notes > No. 1 Date 2017. 3. 21 Version 0.
USA-Federal Communications Commission (FCC) This device complies with part 15 of the FCC Rules. Operation is subject to the following two conditions: (1) This device may not cause harmful interference, and (2) this device must accept any interference received, including interference that may cause undesired operation. This equipment has been tested and found to comply with the limits for a Class B digital device, pursuant to Part 15 of FCC Rules.
Welcome Thank you for purchasing LTE tele-health modem (H3G-900). The LTE tele-health modem receive health data from devices such as blood pressure monitor, glucometer, and weight scale The LTE tele-health modem transmit them into a healthcare center through LTE(verizon) mobile network Packaging List 0. Modem Main AC Adapter (5V/3A) Overview This document describes instructions regarding the H3 LTE tele-health modem (H3G-900, H3 System) and the related components.
Available devices - Lifescan Onetouch Ultra/Mini with Bluetooth adapter(BA-110,BA-300) - Weight : 160g - A&D weighting scale (UC-321PBT) - A&D blood pressure monitor (UA-767PBT) - IEM blood pressure monitor (Stabil-O-Graph) - ForaCare’s thermometer (FORA IR20b) - Beijing Choice’s Pulse Oximeter (MD300C318) - Omron blood pressure monitor (HEM-9200T) - Omron weighting scale (HN-290T) - Welch Allyn blood pressure monitor (RPM-BP100) - Welch Allyn weighting scale (RPM-Scale100) - P1ng Pedometer (PB-001) 2.
Figure 1. Power ON Switch Figure 2.
1. LTE tele-health modem configuration 1) Components A. A serial cable provided by H3 System B. Configuration software i. For admin: Serial numbers can be modified. ii. For user: Serial number change is not allowed. 2) Descriptions A. Configuration software for an admin: Serial No. modification is enabled. B. Items i. Open: Open the serial port ii. Setting: Shall be checked for changing LTE tele-health modem configurations iii. Serial No: Serial number of the LTE tele-health modem.
xi. 4G Log: 4G AT-command logs . Save Folder : C:\log xii. Menu : NetworkPairing submenu : Network Pairing Management UI FirmWare submenu : FOTA File Generate Management UI 3) Steps A. Turn off the LTE tele-health modem. B. Connect a serial cable (provided by H3 System) to a LTE tele-health modem and a PC. C. Run the configuration software. D. Select the suitable COM port (You can check the COM port at the device manager.) and click “Open.” E. If you want to modify configuration, then check “Setting.
A. Admin UI Pairing mode B. Network Pairing mode 2) Admin UI Pairing mode A. General process i. Turn on the LTE tele-health modem. ii. When LTE tele-health modem turns on, LTE tele-health modem shows the number of currently paired devices with LEDs for 5 seconds. (Even at the mode “N”) Number of paired devices 0 1 2 3 4 5 SENDING LED Off Green On Green On Green On Red On Red On READING LED Red On Off Green On Red On Off Green On iii. iv. v. vi. vii. viii. ix. x. xi. xii. xiii. xiv.
3. - AMC instructs the patient to make the BT device be at pairing mode. GW's LED indicates the pairing status. 4. After completion, GW’s LED indicates whether the pairing succeeds or fails. 5. GW automatically updates the pairing results into AMC’s DB through DC 6. 7. C. Final 1. 2. SW. AMC checks whether the process is successfully completed through AMC's DB.
3. Measurement and data transmission 1) How to use A. Take a reading, and then the reading is transmitted to the LTE tele-health modem by Bluetooth. B. And the reading is automatically transmitted to the server via mobile networks. C. At server access failure, it tries 9 more times, in total 10 times at most with 1 hour period until success. 2) Clock setup A. Clock of the LTE tele-health modem is automatically set with 4G connection. B.
4. Data collection software 1) Setup A. Default path : C:\Program Files\H3SYSTEM\DataCollection_GPRS\DataCollection_GPRS.exe B. Configuration file name: gprs_config.ini C. Contents of configuration file (gprs_config.ini) i. LOG_PATH=C:\gprs\log <= Log file path ii. DATA_PATH=C:\gprs\data <= Data XML file path iii. PORT=65001 <= TCP/IP port D. Change of configuration i. Stop the corresponding windows service ii. Change the configuration file, gprs_config.ini iii. Re-start the windows service E.
PAGE 14Item Common serial="2222222222" dc="0" battery="1.2" type="glucose" id="ZHZ306CBY" measuretime="20100101001250" con_gwtime="20111221132941" con_devtime="20111221132941" Glucometer result="103" unit="0" model="1" adapter="1" battery="3.0" BP Monitor (AND) Weight Scale Thermometer BP Monitor (IEM) Oximeter sys="139" dia="091" pul="094" battery="2.63" result="066.80" unit="kg" battery="2.32" otemp=”98.42” btemp=”74.