1. Specification
1. Dimension & weight
- Size : 125 x 100 x33 mm
- Weight : 160g
-
2. Communication technology
- Bluetooth : class 2
- Range : 10m
- Profile : SPP (Serial Port Profile)
-
3. Power supply : AC Adapter (5V/3A) or Rechargeable battery (Lithium polymer)
4. Frequency : Dual band (800, 1900MHz)
5. Operation temp : -10℃ to 40℃(14 ℉ to 104 ℉)
6. Antenna : External Quad band antenna
7. Available devices
-
Lifescan Onetouch Ultra/Mini with Bluetooth adapter(BA-110)
- Weight : 160g
-
A&D weighting scale (UC-321PBT)
-
A&D blood pressure monitor (UA-767PBT)
-
IEM blood pressure monitor (Stabil-O-Graph)
-
ForaCare’s thermometer (FORA IR20b)
-
Beijing Choice’s Pulse Oximeter (MD300C318)
2. Gateway installation
1) Preparation for deployment
A. SIM card
i. Unscrew and open the bottom cover.
And insert a SIM card.
B. Gateway configuration
i. Turn off the gateway.
ii. Connect a serial cable (provided by H3
System) to a gateway and a PC.
iii. Run the configuration software.
iv. Turn on the gateway.
v. Configure the gateway with the configuration software. For the details, please
see the section below, “Gateway configuration.”
C. Pairing and testing
i. Pair with devices. Please see “Pairing” section below.
ii. Take readings and confirm if the readings are successfully delivered to a server.
Please see “Measurement and data transmission” section.
Figure 1. SIM card insertion