User Manual

1) After a measurement, the data is transmitted to the gateway by Bluetooth .
And the data is automatically transmitted to the server via mobile networks.
2) Time set-up
Gateway time set-up
a. A gateway time set-up is done at the beginning of the server access.
b. A gateway time set-up is done before disconnecting the server access.
Device time set-up
a. A&D weight scale, A&D thermometer, J&J Glucometer
- Automatic time set-up is taken after sending data by Bluetooth.
b. IEM B/P monitor: Time set-up should be done in the device manually.
3) Reading LED
Starting Bluetooth communication: Reading GREEN led on
End Bluetooth communication: Reading GREEN led off
At the success of Bluetooth communication: Reading LED blinks GREEN two times.
At the failure of Bluetooth communication: Reading LED blinks RED five times.
4) Data transmission to the server
Sending LED
a. Starting server communication : Sending LED in GREEN is on.
b. End server communication: Sending LED in GREEN is off.
c. At the success of server communication: Sending LED blinks GREEN two times.
d. At the failure of server communication: Sending LED blinks RED five times.
Notification: Server access is taken at power on and power off.
Heart beat: Server access is taken every six hours, after power-on.
At data receiving from device, automatic server access is taken and data is
transmitted.
At the failure of server access, reconnection to the server is done periodically.
Period is every hour. (Max. 3 times)
5. XLM File
<?xml version="1.0" encoding="utf-8" ?>
<Measure serial="1111111111" dc="0" battery="1.2">
<Device type="glucose">
<Data id="ZHZ306CBY" measuretime="20100101001250" result="103" unit="0" model="1" adapter="1" />
</Device>
<Device type="glucose">
<Data id=
"ZHZ306CBY" measuretime="20100101001406" result="112" unit="0" model="1" adapter="1" />
</Device>
<Device type="pressure">