User Manual

1) After a measurement, the data is transmitted to the gateway by Bluetooth .
And the data is automatically transmitted to the server via mobile networks.
2) Time set-up
① Gateway time set-up
a. A gateway time set-up is done at the beginning of the server access.
b. A gateway time set-up is done before disconnecting the server access.
② Device time set-up
a. A&D weight scale, A&D thermometer, J&J Glucometer
- Automatic time set-up is taken after sending data by Bluetooth.
b. IEM B/P monitor: Time set-up should be done in the device manually.
3) Reading LED
① Starting Bluetooth communication: Reading GREEN led on
② End Bluetooth communication: Reading GREEN led off
③ At the success of Bluetooth communication: Reading LED blinks GREEN two times.
④ At the failure of Bluetooth communication: Reading LED blinks RED five times.
4) Data transmission to the server
① Sending LED
a. Starting server communication : Sending LED in GREEN is on.
b. End server communication: Sending LED in GREEN is off.
c. At the success of server communication: Sending LED blinks GREEN two times.
d. At the failure of server communication: Sending LED blinks RED five times.
② Notification: Server access is taken at power on and power off.
③ Heart beat: Server access is taken every six hours, after power-on.
④ At data receiving from device, automatic server access is taken and data is
transmitted.
⑤ At the failure of server access, reconnection to the server is done periodically.
Period is every hour. (Max. 3 times)
5. XLM File
<?xml version="1.0" encoding="utf-8" ?>
<Measure serial="1111111111" dc="0" battery="1.2">
<Device type="glucose">
<Data id="ZHZ306CBY" measuretime="20100101001250" result="103" unit="0" model="1" adapter="1" />
</Device>
<Device type="glucose">
<Data id=
"ZHZ306CBY" measuretime="20100101001406" result="112" unit="0" model="1" adapter="1" />
</Device>
<Device type="pressure">