User Manual

1. Gateway Installation
1) Unscrew and open the bottom cover. And
insert a simcard.
2) To configure, connect a serial cable between a
gateway and a PC and run the configuration
software. And power on the gateway.
Start command is transmitted from a gateway
<Picture1: Simcard Insert>
to configuration software.
If there is any expected response from PC, communication for set-up mode is on.
Otherwise, regular log is transmitted from the gateway to the configuration software.
a. Check ‘setting’ in the configuration software: The set-up mode is activated and
log is also transmitted to the configuration software.
b. Uncheck ‘setting’ in the configuration software: Only log transmission is activated.
2. Pairing
1) Power off.
2) Select the pairing switch and power on.
N: Normal mode (No pairing)
P1: Pairing as a slave. Used for A&D weight
scale.
P2: Pairing as a master. Used for IEM B/P
monitor, J&J glucometer and ForaCare thermometer.
<Picture3: Pairing Switch>
* After pairing is done, it turns to normal operation mode.
2 ) LED
On pairing, Reading LED in GREEN is on.
At the success of pairing, Reading LED blinks GREEN two times.
At the failure of pairing, Reading LED blinks RED five times.
4) Device pairing method
A&D weight scale: Select P1 and measure a weight with the device
IEM B/P monitor: Select P2 and change into the pairing mode
J&J glucometer: Select P2 and change into the pairing mode
FORA thermometer: Select P2 and measure a temperature with the device