User's Manual
10
5. [Low battery sign and battery replacement]
If the LED keeps blinking Red just after taking a reading or pressing the Red button,
please replace the batteries with brand-new ones.
Cautions
Be sure to carefully read the user’s manual before use.
Keep the Bluetooth adapter in a clean and dry place away from dust or corrosive gases
as well as away from fires
Do not apply any shock or vibration to the product
Do not rub the product with inflammable such as alcohol, benzene, or acetone the can
cause chemical change. Use a dry cloth to prevent the surface from being damaged.
Do not stretch nor bend the cable by force.
Disassembling the product is not advisable.
Keep a Bluetooth adapter connected to a glucometer. Frequent connection and
disconnection may cause damages on the connector port.
If Bluetooth communication failure continues for a long time, the oldest readings may
not be transferred to the target device. The maximum storage of the adapter is 500
readings.
Troubleshooting
If you have some troubles in data transmission, please check the below
Try to do “Data transfer” by pressing the Red button while a glucometer and the
adapter are being connected.
Replace batteries with new ones.
Disconnect the glucometer, remove batteries, press the Red button several times,
wait for a while, insert batteries again, and connect the glucometer again. And take
a reading and confirm data transmission.