User manual
12
4.ReleasetheEnterkey.
5.Bypassthevoltagemeasuringinput(U=0V)andusetheEnterkey.
6.Connectaconstantvoltagesourcewithapprox.20Vtothevoltagemeasuringinput.
This voltage must also be connected to a reference measuring device simultaneously.
ThedisplayoftheHB624isnowindicatingavaluethatshouldalsobeindicatedbythe
voltmeter.Inthiscase:20000mV.Ifthisvaluediffers,itwillbesetwiththeUp-Down
keys.AfterwardstheadjustmentwillbeconrmedbytheEnterkey.
7.Releaseorbypassthecurentmeasuringinput(I=0mA)andconrmwiththeEnter
key.
Adjustment curent measuring input:
8.Onthecurrentmeasuringinputconnectaconstantcurrentsourceapprox.2A(e.g.a
powersupplyunitwithcurrentlimiting).Thiscurrentshouldalsobecheckedwitharefe-
rencemeasurementdevice.ThedisplayoftheHB624indicatesavaluewhichshould
nowbeindicatedbytheamperemeter.Inthiscase:2000mA.Ifthisvaluediffers,itcan
besetbytheUp-Downkey.AfterwardstheadjustmentwillbeconrmedwiththeEnter
key.