17 EMC INFORMATION
Guidance and manufacture’s declaration – electromagnetic emission
TC200 is intended for use in the electromagnetic environment specified below. The customer or the user of TC200 should assure that
they are used in such an environment.
Emission test
Compliance
Electromagnetic environment – guidance
RF emissions
CISPR 11
Group 1
TC200 uses RF energy only for their internal function. Therefore, their RF emissions
are very low and are not likely to cause any interference in nearby electronic device.
RF emission
CISPR 11
Class B
TC200 is suitable for use in all establishments, other than domestic establishments
and those directly connected to the public low-voltage power supply network that
supplies buildings used for domestic purposes.
Harmonic emissions
IEC/EN 61000-3-2
N/A
Voltage fluctuations/
flicker emissions
IEC/EN 61000-3-3
N/A
NOTE The EMISSIONS characteristics of TC200 make it suitable for use in industrial areas and hospitals (CISPR 11 class A).
If it is used in a residential environment (for which CISPR 11 class B is normally required) TC200 might not offer
adequate protection to radio-frequency communication services. The user might need to take mitigation measures,
such as relocating or re-orienting the device.