User manual

3
2-2. Constant Voltage Operation
(1) Output Voltage ranges from 0 to rating voltage with continuous adjustment.
(2) Voltage regulation
line regulation5mV.
load regulation5mV.
(3) Recovery time500μs(50% Load change, minimum load 0.5A).
(4) Ripple & Noise5mVrms, 100mVp-p (tested by 20MHz oscilloscope.)
(5) Temperature coefficient100ppm/.
2-3. Constant Current Operation
(1) Output current ranges from 0 to rating current with continuous adjustment.
(2) Current regulation
line regulation3mA.
load regulation3mA.
SPS-606 SPS-3610 SPS-2415 SPS-1820 SPS-1230
(3) Ripple & Noise
3mArms 5mArms 10mArms 10mArms 30mArms
2-4. Indicator Meter
1)Voltage:
Display : 3 1/2 Digits 0.39” Green LED display.
Accuracy : ±(0.5% of rdg + 2 digits).
2)Current:
Display : 3 1/2 Digits 0.39” Red LED display.
Accuracy : ±(0.5% of rdg + 2 digits).
2-5. Over Voltage Protection
(1) Over Voltage Protection ranges from 5% rating to rating +5.5%.
(2) OVP Accuracy ±(Vset 1%+0.6V)
2-6. Insulation
Between Chassis and Output Terminal
: 20MΩ (DC500V).
Between Chassis and AC Cord : 30MΩ (DC500V).