(4) Output Test lead Selection:
The Selection of Output Test Lead and Feedback Test Lead:
For safety assurance, please select the adequate output test lead according to the following list:
Conductor UL
(CSA)
Model
Wire No.
AWG
Component
pc/mm
Cross Section Area
(mm)
2
Outer
Diameter
mm
Maximum
Conductive
Resistor
Ω/km
Permissible
Current
A(amp)
24 11/0.16 0.22 0.64 88.6 7.64
22 17/0.16 0.34 0.78 62.5 10.0
20 21/0.18 0.53 0.95 39.5 13.1
18 34/0.18 0.87 1.21 24.4 17.2
16 26/0.254 1.32 1.53 15.6 22.6
14 41/0.254 2.08 2.03 9.90 30.4
12 65/0.254 3.29 2.35 6.24 40.6
1015
TEW
(Twisted Wire)
10 65/0.32 5.23 3.00 3.90 55.3
Remark:
1. The ambient temperature of “Permissible Current” is at 40, the withstanding temperature of conductor is at 105
according to the condition of the distributed single wire.
2. The permissible current listed as above is suggested to be used under 70%.
3. If the feedback test leads are in need, the level above UL(CSA) AWG24, 22, 20… can be accepted. Besides,
when the load is a capacitive load, please use the twine wire by twisting (+)output test lead with (S+) feedback
test lead. Same way used on (-) output test lead and (S-) feedback test lead.
4. When the current value exceeds above suggestive list, can select more wires used in parallel according to above
list.