User Manual
APPLICATIONS
281
Background
The user-defined mask files can be created out of
support (from an external computer for example)
and uploaded to the MSO-2000E/2000EA Mask
application with a USB key device.
Create an unformatted text file respecting the
format described below.
File extension
File_name.MSK
Format
Format,2.0E,
Total Area Number,1,
Area Number,1,
Points Number,3,
0.00,2.00,
1.00,1.00,
-1.00,1.00,
Example
(with Division
units)
Format,2.0E,
Total Area Number,2,
Area Number,1,
Points Number,4,
0.00,2.00,
1.00,1.00,
0.00,0.00,
-1.00,1.00,
Area Number,2,
Points Number,3,
0.00,-2.00,
1.00,-1.00,
-1.00,-1.00,