User Manual

APPLICATIONS
265
To output the Go-NoGo results to an
external device, the Go-NoGo rear
panel terminal (open collector) can be
used. The Go-NoGo terminal will
output a positive pulse each time a
NoGo violation has occurred for a
minimum of 500us. The voltage of the
pulse depends on the external pull-up
voltage.
Enable
NoGo
500
us
Waveform
Go-NoGo signal
220
Go-NoGo
BNC
B
+
CMOS
(Max 16V)
Or TTL
(Max 5V)