User Manual

ADVANCED CONFIGURATION
161
Setup Holdoff Level
The holdoff function defines the waiting period
before the MSO-2000E/2000EA starts triggering
again after a trigger point. The holdoff function
ensures a stable display if there are a number of
points in a periodic waveform that can be
triggered. Holdoff applies to all the triggering
types except the trigger by bus.
Wait for
trigger
Holdoff time
Trigger point
Next trigger point
1. Press the trigger Menu key.
Menu
2. To set the Holdoff time, press the
Holdoff (or Mode/Holdoff) menu
button on the bottom bezel.
3. Use the side menu to set the
Holdoff time.
Range
4ns~10s
Pressing Set to Minimum sets the
Holdoff time to the minimum, 4ns.
Note: The holdoff function is automatically
disabled when the waveform update mode is in
roll mode (page 110).