User Manual
    APPLICATIONS 
265 
Using the Go-
NoGo Output 
To output the Go-NoGo results to an 
external device, the Go-NoGo rear 
panel terminal (open collector) can be 
used. The Go-NoGo terminal will 
output a positive pulse each time a 
NoGo violation has occurred for a 
minimum of 500us. The voltage of the 
pulse depends on the external pull-up 
voltage. 
Timing Diagram 
Enable
NoGo
500
us
Waveform 
Circuit Diagram 
Go-NoGo signal
220
Ω
Go-NoGo 
BNC
B
+ 
CMOS 
(Max 16V) 
Or TTL 
(Max 5V)