User Manual
    MSO-2000E & 2000EA User Manual 
168 
9.  Set the threshold level using the 
side menu. 
Analog 
channel 
Range 
00.0V~ 5 screen divisions 
Set to TTL 1.4V 
Set to ECL -1.3V 
Set to 50% 
Digital 
channel 
Range 
-5.00V~ +5.00V 
TTL    1.4V 
5.0V CMOS  2.5V 
3.3V CMOS  1.65V 
2.5V CMOS  1.25V 
ECL    -1.3V 
PECL    3.7V 
0V    0V 
 Note 
Setting the trigger threshold for a digital source 
will also change the threshold levels set in the 
Logic Analyzer menu (page 217). 