ADVANCED CONFIGURATION
153
7. Press Number of Bytes from the side
menu and choose the number of
bytes for the data.
UART
1~10 Bytes
8. Press Data from the side menu to
edit the triggering data.
To edit the data, use the Variable
knob to highlight a binary or hex
digit and press Select. Use the
Variable knob to choose a value for
the digit and press Select to
confirm.
VARIABLE
Binary
0,1,X (don’t care)
Hex
0~F, X (don’t care)
ASCII
ASCII characters for the equivalent
Hex characters 00 to FF
I
2
C Bus Trigger Settings
The I
2
C bus trigger conditions can be set at any time after the bus
settings has been set to I
2
C.
Panel Operation
1. Set the Bus to I
2
C in the bus menu.
Page 118
2. Press the Trigger Menu key.
Menu
3. Press Type from the bottom menu.
4. Press Others from the side
menu and select Bus.