User Manual

LIST (MULTI STEP MODE)
95
Setting delay
Set the delay time before each test step. The delay time range is
0ms~5000ms.
Steps
1. Press the Measure button to enter
[MEAS DISPLAY] page, and then
press the LIST SET function button
on the right side of the LCD to enter
the [LIST SET] page.
2. Use arrow keys to move cursor and
select DELAY item on the [LIST SET]
page.
3. Use the numeric keypad to enter the delay
time.