Setting output impedance
The output impedance can be set to 2 or 100Ω.
The varied signal source output impedance will lead to the varied
current or the difference of measuring value. If selecting <25Ω>,
then the voltage range is 10mV~1Vrms and the current range is
400µA~40mArms.
If need to compare test results with Keysight select 100Ω.
Steps
1. Press the Measure button to enter
[MEAS DISPLAY] page, and then
press the SWEEP function button on
the right side of the LCD to enter the
[SWEEP DISPLAY] page, and press
the Setup button to enter [SWEEP
MODE SETUP] page.
2. Use arrow keys to move the cursor
and select OUTPUT IMPEDANCE
item on the [SWEEP MODE SETUP]
page.
3. Use the numeric keypad to enter the time (ms).
Available
parameter
100Ω
Set output impedance to 100Ω
25Ω
Set output impedance to 2