User Manual

COMMAND OVERVIEW
217
:SWEep:TRACB:DIVision <value NR3/disc>
:SWEep:TRACB:DIVision?
Function It sets or queries the DIV of TRACE B.
Description
Set parameter The numeric value can only be set to 1, 2, 5 and
10(when Y-axis setup is linear mode).
Set syntax
:SWEep:TRACB:DIVision 2k
Query syntax
:SWEep:TRACB:DIVision?
Return
data 2.000000E+03 (Format is in <NR3>)
:SWEep:TRACB:DECade {1~12}
:SWEep:TRACB:DECade?
Function It sets or queries the DECADE of TRACE B.
Description
Set parameter The value of DECADE is {1~12} (when Y-axis setup is
LOG mode).
Set syntax
:SWEep:TRACB:DECade 5
Query syntax
:SWEep:TRACB:DECade?
Return
data 5 (Format is in <NR3>).
:SWEep:TRACB:MAXimun?
Function It queries the maximum of TRACE B and corresponding
frequency | voltage | current.
Description
Query syntax
:SWEep:TRACB:MAXimun?
Return
data +2.230924E+06,+3.221517E-03 (Format is in
<NR3>).The first data is the corresponding frequency | voltage |
current. The second data is the maximum.