User Manual
COMMAND OVERVIEW
209
:SWEep:VOLTage <voltage NR3/disc>
:SWEep:VOLTage?
Function: It sets or queries the LEVEL to voltage mode, when the
TYPE setup is FREQ.
Description:
Set parameter The value of
voltage AC
RO 100Ω is
{0.01~2|MAXimum|MINimum}.
The value of
voltage AC
RO 25Ω is
{0.01~1|MAXimum|MINimum}.
Set syntax
:SWEep:VOLTage 1
:SWEep:VOLTage 1M
:SWEep:VOLTage 1MV
:SWEep:VOLTage 1E-3
:SWEep:VOLTage MAXimum
:SWEep:VOLTage MINimum
Query syntax
:SWEep:VOLTage?
Return data 1.000000E+03(Format is in <NR3>)
When the LEVEL setup is not voltage mode, the data returns
9.9E37.
:
SWEep:CURRent
<currentNR3/disc>
:SWEep:CURRent?
Function: It sets or queries the LEVEL to current mode, when the
TYPE setup is FREQ.
Description:
Set parameter The value of
current AC
RO 100Ω is
{
0.0002~0.02
|MAXimum|MINimum}.
The value of
current AC
RO 25Ω is
{
0.0002~0.04
|MAXimum|MINimum}.
Set syntax
:SWEep:CURRent 1
:SWEep:CURRent 1M
:SWEep:CURRent 1MA
:SWEep:CURRent 1E-3