User Manual
COMMAND OVERVIEW
187
:MEASure:RANGe:AC {1|2|3|4|AUTO|HOLD}
Function: It sets or queries the AC RANGE.
Description:
Set parameter {1|2|3|4|
AUTO
|
HOLD
}
Set syntax
:MEASure:RANGe:AC 3
:MEASure:RANGe:AC AUTO
Query syntax
:MEASure:RANGe:AC
Return data 3 (Format is in <NR1>)
:MEASure:OIMPedance {100|25}
:MEASure:OIMPedance?
Function: It sets or queries the RO.
Description:
Query syntax
:MEASure:OIMPedance?
Return
data 100|25 (Format is in <NR1>)
:MEASure:COMParator:PARAmeter {1|2|3|4}
:MEASure:COMParator:PARAmeter?
Function: It sets or queries which parameters number is to be
compared.
Description:
Set syntax
:MEASure:COMParator:PARAmeter 3 (After this
command, every compared setup is at this parameters number.)
Query syntax
:MEASure:COMParator:PARAmeter?
Return
data 3 (Format is in <NR1>)